Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Anaesthesia ; 46(9): 767-70, 1991 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-1928680

RESUMO

We modified a Penlon Nuffield 200 for use in a monoplace hyperbaric oxygen chamber by feeding back the chamber pressure to the reducing valve of the Nuffield 200. This provides a compensating mechanism, allowing the ventilator to deliver adequate tidal volumes at pressures of up to 3 atmospheres. We report the laboratory testing of the ventilator and our experience of ventilating two patients with carbon monoxide poisoning. Although compensation is not complete the modification is adequate for short-term clinical use in patients in whom the airway is compromised but who need hyperbaric oxygen therapy.


Assuntos
Intoxicação por Monóxido de Carbono/terapia , Oxigenoterapia Hiperbárica/instrumentação , Ventiladores Mecânicos , Adulto , Humanos , Masculino , Modelos Estruturais , Respiração/fisiologia , Volume de Ventilação Pulmonar/fisiologia
2.
Br Med J ; 1(6170): 1089, 1979 Apr 21.
Artigo em Inglês | MEDLINE | ID: mdl-444948
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA