1.
J Biomech Eng ; 139(3)2017 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-27996078

RESUMO

We present a new approach to designing three-dimensional, physically realizable porous femoral implants with spatially varying microstructures and effective material properties. We optimize over a simplified design domain to reduce shear stress at the bone-prosthetic interface with a constraint on the bone resorption measured using strain energy. This combination of objective and constraint aims to reduce implant failure and allows a detailed study of the implant designs obtained with a range of microstructure sets and parameters. The microstructure sets are either specified directly or constructed using shape interpolation between a finite number of microstructures optimized for multifunctional characteristics. We demonstrate that designs using varying microstructures outperform designs with a homogeneous microstructure for this femoral implant problem. Further, the choice of microstructure set has an impact on the objective values achieved and on the optimized implant designs. A proof-of-concept metal prototype fabricated via selective laser melting (SLM) demonstrates the manufacturability of designs obtained with our approach.


Assuntos
Fêmur , Desenho de Prótese , Reabsorção Óssea/etiologia , Próteses e Implantes/efeitos adversos , Estresse Mecânico
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...