Pacing Clin Electrophysiol ; 35(12): 1498-504, 2012 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-23035735

RESUMO

INTRODUCTION: Insulation defects with externalized conductors have been reported in the St. Jude Riata(®) family of defibrillation leads (St. Jude Medical, Sylmar, CA, USA). The objective of the Northern Ireland Riata(®) lead screening program was to identify insulation defects and externalized conductors by systematic fluoroscopic and electrical assessment in a prospectively defined cohort of patients. We sought to estimate the prevalence, identify risk factors, and determine the natural history of this abnormality. METHODS: All patients with a Riata(®) lead under follow-up at the Royal Victoria Hospital were invited for fluoroscopic imaging and implantable cardioverter-defibrillator lead parameter checks. Fluoroscopic images were read independently by two cardiologists and the presence of externalized conductors was classified as positive, negative, or borderline. RESULTS: One hundred and sixty-five of 212 patients with a Riata lead were evaluated by fluoroscopy and lead parameter measurements. The mean duration after implantation was 3.98+/-1.43 years. After screening 25 (15%) patients were classified as positive, 137 (83%) negative, and three (1.8%) borderline. Time since implantation (P = 0.001), presence of a single coil lead (P = 0.042), and patient age (P = 0.034) were significantly associated with externalized conductors. The observed rate of externalized conductors was 26.9% for 8-French and 4.7% for 7-French leads. No leads that were identified prospectively with externalized conductors had electrical abnormalities. Seven of 25 (28%) patients had a defective lead extracted by the end of this screening period. CONCLUSION: A significant proportion (15%) of patients with a Riata lead had an insulation breach 4 years after implantation. High-resolution fluoroscopic imaging in at least two orthogonal views is required to identify this abnormality.


Assuntos
Desfibriladores Implantáveis/efeitos adversos , Eletrodos Implantados , Análise de Falha de Equipamento , Falha de Equipamento/estatística & dados numéricos , Fluoroscopia/métodos , Idoso , Distribuição de Qui-Quadrado , Desenho de Equipamento , Segurança de Equipamentos , Feminino , Humanos , Masculino , Pessoa de Meia-Idade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...