Optom Vis Sci ; 71(4): 290-2, 1994 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-8047344

RESUMO

Our study was developed to measure the effect of swabbing on the integrity of Haag-Streit tonometer tips over time. The following solutions were used: hydrogen peroxide, dilute bleach, alcohol, phenol-phenate, and distilled water (as a control). The results show that swabbing causes surface scratching of tonometer tips after 1 year. The results also show that alcohol swabbing will lead to breakdown of the glue and warrant replacement of the tonometer tip after 2 years use.


Assuntos
Desinfetantes/efeitos adversos , Falha de Equipamento , Tonometria Ocular/instrumentação , Humanos , Tampões de Gaze Cirúrgicos