Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Assunto principal
Intervalo de ano de publicação
1.
AMIA Annu Symp Proc ; 2016: 490-495, 2016.
Artigo em Inglês | MEDLINE | ID: mdl-28269844

RESUMO

Interoperability is a major challenge in current healthcare systems. It brings big hope for data exchange, but also raises some concern about patient safety. We study the wireless updating of modern infusion pumps and demonstrate the possible flaws in this process. Through analyzing data on drug limit libraries (DLL) versions in one hospital we could identify the delays in distributing DLL updates and the impact these delays might have on patient safety. We found that 31% of all started infusions had used outdated DLL versions, and 22.6% of all alerts were triggered by outdated DLLs. These findings suggest that clinical and operational stakeholders in healthcare systems must address the unreliable interoperability of medical technologies such as seen on infusion pumps. The impact of information inconsistency across healthcare systems might result in use error which would impair patient safety.


Assuntos
Bombas de Infusão , Segurança de Equipamentos , Humanos , Sistemas de Registro de Ordens Médicas , Erros de Medicação/prevenção & controle , Integração de Sistemas , Tecnologia sem Fio
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...