1.
Opt Lett ; 33(10): 1062-4, 2008 May 15.
Artigo em Inglês | MEDLINE | ID: mdl-18483512

RESUMO

A novel configuration of stationary Fourier transform infrared (FTIR) spectrometer is presented. Contrary to classic configurations, the interferometer is directly integrated in the focal plane array (FPA) during its process of fabrication. A first, to the best of our knowledge, demonstration of the spectrometric function has been achieved departing from a well-known structure of an HgCdTe photodetector. We show that the obtained FTIR-FPA can be described by intrinsic parameters such as an optical path difference and a so-called spectrometric efficiency. First experimental results are presented.