Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Intervalo de ano de publicação
1.
Herald of Medicine ; (12): 650-652, 2014.
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-446229

RESUMO

Objective To establish a RP-HPLC method for the determination of 3-amide-indole derivative and its related substances. Methods Diamonsil C18(250 mmí4. 6 mm,5 μm) column was adopted. The mobile phase consisted of a mixture of methanol-acetonitrile-water(431245) at the flow rate of 1. 0 mL·min-1 . The wavelength for ultraviolet detection was 224 nm. The injection volume was 20 μL and the column temperature was room temperature. Results 3-amide-indole derivative and related substances could be well separated. The linearity of the 3-amide-indole derivative curve was well correlated (r=0. 999 7) within the range of 0. 04-0. 16 mg·mL-1. The RSD was 0. 52%with good repeatability. The detection limit was 2. 65 ng. Conclusion The method is accurate,reliable,sensitive and specific,which could be used for the determination of 3-amide-indolederivative and related substances.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...