RESUMO
In this paper the design restrictions of an inductive sensor for an intraocular lens with focus control on the basis of a marker implanted in the ciliary muscle of the eye are discussed in the framework of anatomical and physiological influences and constraints: limitations on the marker size, influences of tissue conduction and effects of off-axis implantation of the marker with respect to the coil.
Assuntos
Corpo Ciliar/fisiologia , Lentes Intraoculares , Corpo Ciliar/anatomia & histologia , Humanos , Próteses e Implantes , TelemetriaRESUMO
In this paper a concept is proposed of an intraocular lens implant with electro-optic accommodation of a variable-focus hybrid liquid-crystal-based lens. The dioptric strength of the lens is electronically controlled by a signal that is derived from the change of inductance of a sensing coil due to a marker implanted in the nearby contracting or decontracting ciliary muscle. Analytical, numerical and experimental results are reported on the dependency of the frequency of a Colpitts oscillator circuit on the location of a nearby conductive marker. A concept is also reported on the use as an electro-optic lens of a device based on a liquid crystal in planar alignment, which is held between a flat and a curved window coated with optically transparent and electrically conductive layers.