J Med Phys ; 49(1): 103-109, 2024.
Artigo em Inglês | MEDLINE | ID: mdl-38828077

RESUMO

Background: The slice spacing has a crucial role in the accuracy of computed tomography (CT) images in sagittal and coronal planes. However, there is no practical method for measuring the accuracy of the slice spacing. Purpose: This study proposes a novel method to automatically measure the slice spacing using the American Association of Physicists in Medicine (AAPM) CT performance phantom. Methods: The AAPM CT performance phantom module 610-04 was used to measure slice spacing. The process of slice spacing measurement involves a pair of axial images of the module containing ramp aluminum objects located at adjacent slice positions. The middle aluminum plate of each image was automatically segmented. Next, the two segmented images were combined to produce one image with two stair objects. The centroid coordinates of two stair objects were automatically determined. Subsequently, the distance between these two centroids was measured to directly indicate the slice spacing. For comparison, the slice spacing was calculated by accessing the slice position attributes from the DICOM header of both images. The proposed method was tested on phantom images with variations in slice spacing and field of view (FOV). Results: The results showed that the automatic measurement of slice spacing was quite accurate for all variations of slice spacing and FOV, with average differences of 9.0% and 9.3%, respectively. Conclusion: A new automated method for measuring the slice spacing using the AAPM CT phantom was successfully demonstrated and tested for variations of slice spacing and FOV. Slice spacing measurement may be considered an additional parameter to be checked in addition to other established parameters.

2.
Biomed Phys Eng Express ; 10(4)2024 May 22.
Artigo em Inglês | MEDLINE | ID: mdl-38744255

RESUMO

Purpose. To develop a method to extract statistical low-contrast detectability (LCD) and contrast-detail (C-D) curves from clinical patient images.Method. We used the region of air surrounding the patient as an alternative for a homogeneous region within a patient. A simple graphical user interface (GUI) was created to set the initial configuration for region of interest (ROI), ROI size, and minimum detectable contrast (MDC). The process was started by segmenting the air surrounding the patient with a threshold between -980 HU (Hounsfield units) and -1024 HU to get an air mask. The mask was trimmed using the patient center coordinates to avoid distortion from the patient table. It was used to automatically place square ROIs of a predetermined size. The mean pixel values in HU within each ROI were calculated, and the standard deviation (SD) from all the means was obtained. The MDC for a particular target size was generated by multiplying the SD by 3.29. A C-D curve was obtained by iterating this process for the other ROI sizes. This method was applied to the homogeneous area from the uniformity module of an ACR CT phantom to find the correlation between the parameters inside and outside the phantom, for 30 thoracic, 26 abdominal, and 23 head images.Results. The phantom images showed a significant linear correlation between the LCDs obtained from outside and inside the phantom, with R2values of 0.67 and 0.99 for variations in tube currents and tube voltages. This indicated that the air region outside the phantom can act as a surrogate for the homogenous region inside the phantom to obtain the LCD and C-D curves.Conclusion. The C-D curves obtained from outside the ACR CT phantom show a strong linear correlation with those from inside the phantom. The proposed method can also be used to extract the LCD from patient images by using the region of air outside as a surrogate for a region inside the patient.


Assuntos
Algoritmos , Tomografia Computadorizada por Raios X , Humanos , Tomografia Computadorizada por Raios X/métodos , Imagens de Fantasmas , Processamento de Imagem Assistida por Computador/métodos , Interface Usuário-Computador , Interpretação de Imagem Radiográfica Assistida por Computador/métodos
3.
Biomed Phys Eng Express ; 10(2)2024 Feb 23.
Artigo em Inglês | MEDLINE | ID: mdl-38359442

RESUMO

Purpose. The use of the Hough transform for angle detection is quite accurate for relatively wide slice thickness. However, the Hough transform fails to accurately detect the angle for thin slice thickness. This study proposes a method for automatically measuring the thickness of thin slices on images of a Catphan phantom.Methods. In the proposed method, the angle of the phantom's orientation was determined based on the relative coordinates of the four hole objects in the phantom. After the angles of the wires were determined, the profiles of pixel values across the wire objects were constructed. Finally, their full widths at half maximum (FWHMs) were determined and multiplied bytan23° to obtain the slice thicknesses of the images. The results of the proposed method were compared to a previous method, which used the Hough transform to obtain the phantom's orientation. We used slice thicknesses ranging from 0.8 mm to 5.0 mm, and phantom angles from 0° to 10°.Results. Our proposed method detected the angle of the phantom accurately for thin slices, whereas a previous method did not accurately detect the angle. The results of the slice thickness using this current method were slightly higher (within 7.9%) compared to the previous method. However, the results of the two methods did not differ significantly (p-value > 0.05). Using different angles, the current method detected all the angles more accurately. Again, the slice thicknesses were not significantly different from the previous method (p-value > 0.05).Conclusion. The proposed method for measuring the thickness of thin slices in an image of a Catphan phantom, based on the relative coordinates of the four hole objects in the phantom, outperformed a previous method based on the Hough transform.


Assuntos
Tomografia Computadorizada por Raios X , Tomografia Computadorizada por Raios X/métodos , Imagens de Fantasmas
4.
Biomed Phys Eng Express ; 9(6)2023 10 11.
Artigo em Inglês | MEDLINE | ID: mdl-37788647

RESUMO

Purpose. The aim of this study is to develop software to automatically assess the laser alignment on the ACR CT phantom and evaluate its accuracy on sixteen CT scanners.Methods. Software for an automated method of laser alignment assessment on the ACR CT phantom was developed. Laser alignment assessment was based on the positions of the ball-bearing markers at the edge of the ACR CT phantom. The automatic assessment was performed using several steps, including segmentation to acquire the coordinates of the ball-bearing markers and determination of the distances between lines connecting them with lines through the center of the image. A comparison of the results from the automatic method with those from the manual method was performed. The manual measurements were carried out using MicroDicom Viewer. A Mann-Whitney U test was performed to determine the statistical difference between both methods. The evaluation was performed on images of the ACR CT phantom scanned with 16 CT scanners from 5 different CT manufacturers.Results. The results confirmed that our software successfully segments the ball-bearing markers and determines the laser alignment assessment on the ACR CT phantom. Evaluation of the algorithm with images from the 16 CT scanners revealed that the difference between the results from automatic and manual methods were about 0.2 mm with apvalue of around 0.7 (no statistical difference). Misalignment in they-axis was larger than the misalignment in the x-axisfor the majority of the scanners tested. It was found that the phantom tended to be placed 2 mm higher than the iso-center.Conclusions. Software to automatically assess CT laser alignment with the ACR CT phantom was successfully developed and evaluated. The automatic assessment was comparable to manual assessment. In addition, the automatic method was user independent and fast.


Assuntos
Algoritmos , Software , Tomógrafos Computadorizados , Imagens de Fantasmas , Tomografia Computadorizada por Raios X/métodos
5.
J Biomed Phys Eng ; 13(4): 353-362, 2023 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-37609515

RESUMO

Background: Methods for segmentation, i.e., Full-segmentation (FS) and Segmentation-rotation (SR), are proposed for maintaining Computed Tomography (CT) number linearity. However, their effectiveness has not yet been tested against noise. Objective: This study aimed to evaluate the influence of noise on the accuracy of CT number linearity of the FS and SR methods on American College of Radiology (ACR) CT and computational phantoms. Material and Methods: This experimental study utilized two phantoms, ACR CT and computational phantoms. An ACR CT phantom was scanned by a 128-slice CT scanner with various tube currents from 80 to 200 mA to acquire various noises, with other constant parameters. The computational phantom was added by different Gaussian noises between 20 and 120 Hounsfield Units (HU). The CT number linearity was measured by the FS and SR methods, and the accuracy of CT number linearity was computed on two phantoms. Results: The two methods successfully segmented both phantoms at low noise, i.e., less than 60 HU. However, segmentation and measurement of CT number linearity are not accurate on a computational phantom using the FS method for more than 60-HU noise. The SR method is still accurate up to 120 HU of noise. Conclusion: The SR method outperformed the FS method to measure the CT number linearity due to its endurance in extreme noise.

6.
Am J Infect Control ; 51(12): 1314-1320, 2023 12.
Artigo em Inglês | MEDLINE | ID: mdl-37478909

RESUMO

BACKGROUND: We assessed the association between neighborhood area deprivation index (ADI) and community-onset (co) and hospital-onset (ho) Staphylococcus aureus infection. METHODS: Demographic and clinical characteristics of patients admitted to 5 adult hospitals in the mid-Atlantic between 2016 and 2018 were obtained. The association of ADI with methicillin-resistant (MRSA) and methicillin-sensitive (MSSA) S aureus infections was assessed using logistic regression models adjusting for severity of illness and days of admission. RESULTS: Overall, increasing ADI was associated with higher odds of co- and ho-MRSA and MSSA infection. In univariate analysis, Black race was associated with 44% greater odds of ho-MRSA infection (odds ratio [OR] 1.44; 95% CI 1.18-1.76) and Asian race (co-MRSA OR 0.355; Confidence Interval (CI) 0.240-0.525; co-MSSA OR 0.718; CI 0.557-0.928) and unknown race (co-MRSA OR 0.470; CI 0.365-0.606; co-MSSA OR 0.699; CI 0.577-0.848) was associated with lower odds of co-MSSA and co-MRSA infections. When both race and ADI were included in the model, Black race was no longer associated with ho-MRSA infections whereas Asian and unknown race remained associated with lower odds of co-MRSA and co-MSSA infection. In the multivariable logistic regression, ADI was consistently associated with increased odds of S aureus infection (co-MRSA OR 1.132; CI 1.064-1.205; co-MSSA OR 1.089; CI 1.030-1.15; ho-MRSA OR 1.29; CI 1.16-1.43: ho-MSSA OR 1.215; CI 1.096-1.346). CONCLUSIONS: The area deprivation index is associated with community and hospital-onset MRSA and MSSA infections.


Assuntos
Infecção Hospitalar , Staphylococcus aureus Resistente à Meticilina , Infecções Estafilocócicas , Adulto , Humanos , Staphylococcus aureus , Infecções Estafilocócicas/epidemiologia , Meticilina , Infecção Hospitalar/epidemiologia , Fatores de Risco
7.
Med Care ; 61(7): 484-489, 2023 07 01.
Artigo em Inglês | MEDLINE | ID: mdl-37289564

RESUMO

BACKGROUND: Social factors are a key determinant of hospital readmission. We describe the development of the country's first statewide policy providing hospitals with financial incentives to reduce readmission disparities. OBJECTIVE: To describe the development and evaluation of a novel program that measures hospital-level disparity in readmission and rewards hospitals for improvement. RESEARCH DESIGN: Observational study using inpatient claims. PARTICIPANTS: Baseline data included 454,372 all-cause inpatient discharges in 2018 and 2019. Of the included discharges, 34.01% involved Black patients, 40.44% involved female patients, 33.1% involved patients covered by Medicaid, and 11.76% involved patients who were readmitted. Mean age was 55.18. MEASURES: The key measure was the percentage change over time within the hospital in readmission disparity. Readmission disparity was measured using a multilevel model that gauged the association between social factors and readmission risk at a given hospital. Three social factors (Race, Medicaid coverage, and Area Deprivation Index) were combined into an index reflecting exposure to social adversity. RESULTS: Of the State's 45 acute-care hospitals, 26 exhibited improved disparity performance in 2019. LIMITATIONS: The program is limited to inpatients within a single state; the analysis does not provide evidence on the causal relationship between the intervention and readmission disparities. CONCLUSION: This represents the first large-scale effort in the US to link disparities to hospital payment. Because the methodology relies on claims data, it could easily be adopted elsewhere. The incentives are directed to within-hospital disparities, thus mitigating concerns about penalizing hospitals with patients with greater social exposure. This methodology could be used to measure disparity in other outcomes.


Assuntos
Motivação , Readmissão do Paciente , Estados Unidos , Humanos , Feminino , Pessoa de Meia-Idade , Maryland , Medicaid , Hospitais
8.
Biomed Phys Eng Express ; 9(4)2023 05 30.
Artigo em Inglês | MEDLINE | ID: mdl-37216929

RESUMO

Objective. To develop an algorithm to measure slice thickness running on three types of Catphan phantoms with the ability to adapt to any misalignment and rotation of the phantoms.Method. Images of Catphan 500, 504, and 604 phantoms were examined. In addition, images with various slice thicknesses ranging from 1.5 to 10.0 mm, distance to the iso-center and phantom rotations were also examined. The automatic slice thickness algorithm was carried out by processing only objects within a circle having a diameter of half the diameter of the phantom. A segmentation was performed within an inner circle with dynamic thresholds to produce binary images with wire and bead objects within it. Region properties were used to distinguish wire ramps and bead objects. At each identified wire ramp, the angle was detected using the Hough transform. Profile lines were then placed on each ramp based on the centroid coordinates and detected angles, and the full-width at half maximum (FWHM) was determined for the average profile. The slice thickness was obtained by multiplying the FWHM by the tangent of the ramp angle (23°).Results. Automatic measurements work well and have only a small difference (<0.5 mm) from manual measurements. For slice thickness variation, automatic measurement successfully performs segmentation and correctly locates the profile line on all wire ramps. The results show measured slice thicknesses that are close (<3 mm) to the nominal thickness at thin slices, but slightly deviated for thicker slices. There is a strong correlation (R2= 0.873) between automatic and manual measurements. Testing the algorithm at various distances from the iso-center and phantom rotation angle also produced accurate results.Conclusion. An automated algorithm for measuring slice thickness on three types of Catphan CT phantom images has been developed. The algorithm works well on various thicknesses, distances from the iso-center, and phantom rotations.


Assuntos
Algoritmos , Tomografia Computadorizada por Raios X , Imagens de Fantasmas , Tomografia Computadorizada por Raios X/métodos
9.
Biomed Phys Eng Express ; 9(3)2023 04 06.
Artigo em Inglês | MEDLINE | ID: mdl-36990062

RESUMO

This study aims to develop a program in Python language for automatic measurement of slice thickness in computed tomography (CT) images of a Siemens phantom with different values of slice thickness, field of view (FOV), and pitch. A Siemens phantom was scanned using a Siemens 64-slice Somatom Perspective CT scanner with various slice thicknesses (i.e. 2, 4, 6, 8, and 10 mm), FOVs (i.e. 220, 260, and 300 mm), and pitch (i.e. 0.7, 0.9, and 1). Automatic measurement of slice thickness was performed by segmenting the ramp insert in the image and detecting angles of the ramp insert using the Hough transform. The resulting angles were subsequently used to rotate the image. Profiles of pixel along the ramp insert were made from the rotated images, and the slice thickness was calculated by determining the full-width at half maximum (FWHM) of the profiles. The product of the FWHM in pixels and the pixel size was corrected by the tangent of the ramp insert (i.e., 23°) to obtain the measured slice thickness. The results of the automatic measurements were compared with manual measurements carried out using a MicroDicom Viewer. The differences between the automatic and manual measurements at all slice thicknesses were less than 0.30 mm. The automatic and manual measurements had high linear correlations. For variations of the FOV and pitch, the differences between the automatic and manual measurement were less than 0.16 mm. The automatic and manual measurements were significantly different (p-value < 0.05) for slice thickness variation. In addition, the automatic and manual measurements were not significantly different (p-value > 0.05) for variations of FOV and pitch.


Assuntos
Tomografia Computadorizada por Raios X , Tomografia Computadorizada por Raios X/métodos , Tomógrafos Computadorizados , Imagens de Fantasmas
10.
Biomed Phys Eng Express ; 9(2)2023 02 03.
Artigo em Inglês | MEDLINE | ID: mdl-36689768

RESUMO

Radiation of thoracic computed tomography (CT) involves the breast although it is not considered an organ of interest. According to the International Commission on Radiological Protection (ICRP) No. 103, the breast is an organ with a high level of sensitivity when interacting with x-rays, increasing the potential risk of breast cancer. Therefore, the radiation dose must be optimized while maintaining image quality. The dose optimization can be accomplished using a radiation shield. This study aims to determine the effect of silicone rubber (SR)-lead (Pb) in various thicknesses as an alternative protective material limiting dose and preserving the image quality of the breast in thoracic CT. SR-Pb was made from SR and Pb by a simple method. The SR-Pb had thicknesses of 3, 6, 9, and 12 mm. The breast dose was measured using a CT dose profiler on the surface of the breast phantom. The CT number and the noise level of the resulting image were determined quantitatively. The dose without the radiation shield was 5.4 mGy. The doses measured using shielding with thicknesses of 3, 6, 9, and 12 mm were 5.2, 4.5, 4.3, and 3.3 mGy, respectively. Radiation shielding with a thickness of 12 mm reduced breast surface dose by up to 38%. The CT numbers and noise levels for the left and right breast phantom images were almost the same as those ​​without radiation shields indicating there were only slight artifacts in the image. Therefore, SR-Pb is considered a good shielding material which can be pplied in a clinical setting by placing it directly on the breast surface for dose optimization.


Assuntos
Chumbo , Elastômeros de Silicone , Doses de Radiação , Bismuto , Tomografia Computadorizada por Raios X/métodos
11.
Appl Radiat Isot ; 192: 110605, 2023 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-36502735

RESUMO

The aim of this study is to measure the volumetric computed tomography dose index (CTDIvol) for different tube voltages for a polyester-resin (PESR) phantom, and to compare it to values for a standard polymethyl methacrylate (PMMA) phantom. Both phantoms are head phantoms with a diameter of 16 cm. The phantoms were scanned by a CT scanner (GE Revolution EVO 64/128 slice) with tube voltages of 80, 100, 120, and 140 kV. The other scan parameters were constant (i.e. tube current of 100 mA, rotation time of 1 s, and collimation width of 10 mm). The CTDI100,c and CTDI100,p were obtained by measuring the dose with an ionization chamber inserted into five holes within the phantoms. The CTDIvol was calculated based on the CTDI100,c and CTDI100,p values. The measurements were repeated three times for each hole. It was found that the CTDIvol values for the PESR phantom were dependent on tube voltage value, and were similar to the dependency in a PMMA phantom. The maximum CTDIvol difference between the PESR and PMMA phantoms was 7.5%. We conclude that the dose measured in the PESR phantom is similar to that in the PMMA phantom and that the PESR phantom can be used as an alternative if the PMMA phantom is not available.


Assuntos
Polimetil Metacrilato , Tomografia Computadorizada por Raios X , Doses de Radiação , Tomografia Computadorizada por Raios X/métodos , Método de Monte Carlo , Tomógrafos Computadorizados , Imagens de Fantasmas
12.
Biomed Phys Eng Express ; 9(1)2022 12 16.
Artigo em Inglês | MEDLINE | ID: mdl-36541467

RESUMO

We developed a software to automatically measure the linearity between the CT numbers and densities of objects using an ACR 464 CT phantom, and investigated the CT number linearity of 16 different CT scanners. The software included a segmentation-rotation method. After segmenting five objects within the phantom image, the software computed the mean CT number of each object and plotted a graph between the CT numbers and densities of the objects. Linear regression and coefficients of regression, R2, were automatically calculated. The software was used to investigate the CT number linearity of 16 CT scanners from Toshiba, Siemens, Hitachi, and GE installed at 16 hospitals in Indonesia. The linearity of the CT number obtained on most of the scanners showed a strong linear correlation (R2> 0.99) between the CT numbers and densities of the five phantom materials. Two scanners (Siemens Emotion 16) had the strongest linear correlation withR2= 0.999, and two Hitachi Eclos scanners had the weakest linear correlation withR2< 0.99.


Assuntos
Acreditação , Software , Tomógrafos Computadorizados , Imagens de Fantasmas , Tomografia Computadorizada por Raios X
13.
J Appl Clin Med Phys ; 23(9): e13719, 2022 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-35808971

RESUMO

PURPOSE: We have developed a software to automatically find the contrast-detail (C-D) curve based on the statistical low-contrast detectability (LCD) in images of computed tomography (CT) phantoms at multiple cell sizes and to generate minimum detectable contrast (MDC) characteristics. METHODS: A simple graphical user interface was developed to set the initial parameters needed to create multiple grid region of interest of various cell sizes with a 2-pixel increment. For each cell in the grid, the average CT number was calculated to obtain the standard deviation (SD). Detectability was then calculated by multiplying the SD of the mean CT numbers by 3.29. This process was automatically repeated as many times as the cell size was set at initialization. Based on the obtained LCD, the C-D curve was obtained and the target size at an MDC of 0.6% (i.e., 6-HU difference) was determined. We subsequently investigated the consistency of the target sizes for a 0.6% MDC at four locations within the homogeneous image. We applied the software to images with six noise levels, images of two modules of the American College of Radiology CT phantom, images of four different phantoms, and images of four different CT scanners. We compared the target sizes at a 0.6% MDC based on the statistical LCD and the results from a human observer. RESULTS: The developed system was able to measure C-D curves from different phantoms and scanners. We found that the C-D curves follow a power-law fit. We found that higher noise levels resulted in a higher MDC for a target of the same size. The low-contrast module image had a slightly higher MDC than the distance module image. The minimum size of an object detected by visual observation was slightly larger than the size using statistical LCD. CONCLUSIONS: The statistical LCD measurement method can generate a C-D curve automatically, quickly, and objectively.


Assuntos
Software , Tomografia Computadorizada por Raios X , Algoritmos , Humanos , Imagens de Fantasmas , Doses de Radiação , Tomógrafos Computadorizados , Tomografia Computadorizada por Raios X/métodos
14.
Biomed Phys Eng Express ; 8(4)2022 05 24.
Artigo em Inglês | MEDLINE | ID: mdl-35550574

RESUMO

A bolus is a material equivalent to soft tissue and is directly placed on the skin surface during radiotherapy. It is commonly used to increase the dose on the skin surface in electron beam radiation. A typical material for a bolus is silicone rubber (SR). We made a bolus with dimensions of 17 × 17 × 1 cm3by varying silicone rubber (SR) RTV 52 and hardening material (bluesil catalyst 60 R) using a simple molded method. We characterized it using a CT scan to find the relative electron density (RED) and examined it using the electron beam of a linear accelerator (LINAC) at energies of 5 and 7 MeV to investigate the percentage of surface dose (PSD). The PSD value is relative to the dose at maximum doses (dmax). The RED value of the bolus was from 1.168 ± 0.021 to 1.176 ± 0.019, higher than the soft tissue (muscle) value of 1.043. The percentage of surface dose (PSD) test at 5 and 7 MeV LINAC energy showed that the highest PSD without using a bolus were 84.79±0.06% and 86.03±0.07%, respectively. With a bolus, the PSD values were 112.52±0.16% and 111.14±0.03%, respectively. The results indicate that bolus fabrication using SR RTV 52 and bluesil 60R is very effective for radiotherapy in the treatment of skin cancer due to an increase in surface dose.


Assuntos
Elétrons , Elastômeros de Silicone , Aceleradores de Partículas , Imagens de Fantasmas , Dosagem Radioterapêutica
15.
Am J Infect Control ; 50(12): 1352-1354, 2022 12.
Artigo em Inglês | MEDLINE | ID: mdl-35217092

RESUMO

BACKGROUND: Previous single-center studies suggest that exposure to a room previously occupied by a patient with CDI may increase the risk of CDI in subsequent patients. We evaluated the risk of previous room occupant on CDI risk across 5 adult hospitals. METHODS: This is a non-concurrent cohort study of adult inpatients admitted to 5 hospitals. Exposed rooms were identified as being occupied by a patient diagnosed with CDI and a logistic regression was performed to assess if staying in an exposed room increases the risk of CDI in subsequent patients. RESULTS: Patients admitted to a room that was previously occupied by a patient with CDI had a 27% increased odds of subsequently being diagnosed with CDI (odds ratio (OR)=1.269; 95% confidence interval (CI)= 1.12-1.44) if exposed within the last 90 days and 40% increased odds (OR=1.401; 95% CI= 1.25-1.57) if exposed in the last 365 days after controlling for previous admissions and length of stay. Cumulative patient-day exposure to previously CDI-positive occupied rooms within both 90 and 365 days were also found to be independently significant, with a 4.5% (OR 1.045; 95% CI = 1.03-1.06) and 4.2% (OR 1.042; 95% CI = 1.03-1.06) increase in odds of CDI with each day of exposure respectively. DISCUSSION/CONCLUSIONS: This study adds further evidence that hospital environment in patient rooms may contribute to risk for CDI.


Assuntos
Clostridioides difficile , Infecções por Clostridium , Infecção Hospitalar , Adulto , Humanos , Estudos de Coortes , Infecção Hospitalar/epidemiologia , Infecções por Clostridium/epidemiologia , Hospitais , Fatores de Risco , Estudos Retrospectivos
16.
Am J Cardiol ; 165: 58-64, 2022 02 15.
Artigo em Inglês | MEDLINE | ID: mdl-34906366

RESUMO

Given the concern that beta-blocker use may be associated with an increased risk for heart failure (HF) in populations with normal left ventricular systolic function, we evaluated the association between beta-blocker use and incident HF events, as well as loop diuretic initiation in the Systolic Blood Pressure Intervention Trial (SPRINT). SPRINT demonstrated that a blood pressure target of <120 mm Hg reduced cardiovascular outcomes compared with <140 mm Hg in adults with at least one cardiovascular risk factor and without HF. The lower rate of the composite primary outcome in the 120 mm Hg group was primarily driven by a reduction in HF events. Subjects on a beta blocker for the entire trial duration were compared with subjects who never received a beta blocker after 1:1 propensity score matching. A competing risk survival analysis by beta-blocker status was performed to estimate the effect of the drug on incident HF and was then repeated for a secondary end point of cardiovascular disease death. Among the 3,284 propensity score-matched subjects, beta-blocker exposure was associated with an increased HF risk (hazard ratio 5.86; 95% confidence interval 2.73 to 13.04; p <0.001). A sensitivity analysis of propensity score-matched cohorts with a history of coronary artery disease or atrial fibrillation revealed the same association (hazard ratio 3.49; 95% confidence interval 1.15 to 10.06; p = 0.028). In conclusion, beta-blocker exposure in this secondary analysis was associated with increased incident HF in subjects with hypertension without HF at baseline.


Assuntos
Antagonistas Adrenérgicos beta/uso terapêutico , Insuficiência Cardíaca/epidemiologia , Hipertensão/tratamento farmacológico , Idoso , Anti-Hipertensivos/uso terapêutico , Fibrilação Atrial/epidemiologia , Pressão Sanguínea , Doença da Artéria Coronariana/epidemiologia , Feminino , Humanos , Hipertensão/epidemiologia , Incidência , Masculino , Pessoa de Meia-Idade , Planejamento de Assistência ao Paciente , Pontuação de Propensão , Modelos de Riscos Proporcionais , Fatores de Risco , Inibidores de Simportadores de Cloreto de Sódio e Potássio/uso terapêutico
17.
J Biomed Phys Eng ; 11(5): 563-572, 2021 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-34722401

RESUMO

BACKGROUND: Estimation of eye lens dose is important in head computed tomography (CT) examination since the eye lens is a sensitive organ to ionizing radiation. OBJECTIVE: The purpose of this study is to compare estimations of eye lens dose in head CT examinations using local size-specific dose estimate (SSDE) based on size-conversion factors of the American Association of Physicists in Medicine (AAPM) Report No. 293 with those based on size-conversion factors of the AAPM Report No. 220. MATERIAL AND METHODS: This experimental study is conducted on a group of patients who had undergone nasopharyngeal CT examination. Due to the longitudinal (z-axis) dose fluctuation, the average global SSDE and average local SSDE (i.e. particular slices where the eyes are located) were investigated. All estimates were compared to the measurement results using thermo-luminescent dosimeters (TLDs). The estimated and measured doses were implemented for 14 patients undergoing nasopharyngeal CT examination. RESULTS: It was found that the percentage differences of the volume CT dose index (CTDIvol), average global SSDE based on AAPM No. 220 (SSDEo,g), average local SSDE based on AAPM No. 220 (SSDEo,l), average global SSDE based on AAPM No. 293 (SSDEn,g) and average local SSDE based on AAPM No. 293 (SSDEn,l) against the measured TLD doses were 22.5, 21.7, 15.0, 9.3, and 2.1%, respectively. All comparisons between dose estimates and TLD measurements gave p-values less than 0.001, except for SSDEn,l (p-value = 0.566). CONCLUSION: SSDE based on AAPM Report No. 293 can be used to accurately estimate eye lens radiation doses by performing the calculations on a number of specific slices containing the eyes.

18.
Radiat Prot Dosimetry ; 196(3-4): 248-256, 2021 Nov 12.
Artigo em Inglês | MEDLINE | ID: mdl-34580737

RESUMO

The study aims to correlate the effective diameter (Deff) and water-equivalent diameter (Dw) parameters with anterior-posterior (AP), lateral (LAT) and AP + LAT dimensions in order to estimate the patient dose in head CT examinations. Seventy-four patient datasets from head CT examinations were retrospectively collected. The patient's sizes were calculated from the middle slice using a software of IndoseCT. Dw and Deff were plotted as functions of AP, LAT and AP + LAT dimensions. The best trendline fit for LAT and AP functions was a second order polynomial, which resulted in R2 of 0.89 for Deff vs LAT, 0.88 for Dw vs LAT, 0.92 for Deff vs AP and 0.91 for Dw vs AP. A linear correlation was found for Deff vs AP + LAT, Dw vs AP + LAT and Dw vs Deff with R2 of 0.97, 0.96 and 0.98, respectively.


Assuntos
Cabeça , Água , Cabeça/diagnóstico por imagem , Humanos , Doses de Radiação , Estudos Retrospectivos , Tomografia Computadorizada por Raios X
19.
J Appl Clin Med Phys ; 22(9): 313-323, 2021 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-34291861

RESUMO

PURPOSE: The aim of this study is to propose an algorithm for the automated calculation of water-equivalent diameter (Dw ) and size-specific dose estimation (SSDE) from clinical computed tomography (CT) images containing one or more substantial body part. METHODS: All CT datasets were retrospectively acquired by the Toshiba Aquilion 128 CT scanner. The proposed algorithm consisted of a contouring stage for the Dw calculation, carried out by taking the six largest objects in the cross-sectional image of the patient's body, followed by the removal of the CT table depending on the center position (y-axis) of each object. Validation of the proposed algorithm used images of patients who had undergone chest examination with both arms raised up, one arm placed down and both arms placed down, images of the pelvic region consisting of one substantial object, and images of the lower extremities consisting of two separated areas. RESULTS: The proposed algorithm gave the same results for Dw and SSDE as the previous algorithm when images consisted of one substantial body part. However, when images consisted of more than one substantial body part, the new algorithm was able to detect all parts of the patient within the image. The Dw values from the proposed algorithm were 9.5%, 15.4%, and 39.6% greater than the previous algorithm for the chest region with one arm placed down, both arms placed down, and images with two legs, respectively. The SSDE values from the proposed algorithm were 8.2%, 11.2%, and 20.6% lower than the previous algorithm for the same images, respectively. CONCLUSIONS: We have presented an improved algorithm for automated calculation of Dw and SSDE. The proposed algorithm is more general and gives accurate results for both Dw and SSDE whether the CT images contain one or more than one substantial body part.


Assuntos
Tomografia Computadorizada por Raios X , Água , Humanos , Pelve , Doses de Radiação , Estudos Retrospectivos
20.
J Appl Clin Med Phys ; 22(7): 313-321, 2021 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-34109738

RESUMO

PURPOSE: The purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso-center, and reconstruction filter. METHODS: An automatic procedure for slice thickness verification on AAPM CT performance phantom was developed using MATLAB R2015b. The stair object image within the phantom was segmented, and the middle stair object was located. Its angle was determined using the Hough transformation, and the image was rotated accordingly. The profile through this object was obtained, and its full-width of half maximum (FWHM) was automatically measured. The FWHM indicated the slice thickness of the image. The automated procedure was applied with variations in three independent parameters, i.e., the slice thickness, the distance from the phantom to the iso-center, and the reconstruction filter. The automated results were compared to manual measurements made using electronic calipers. RESULTS: The differences of the automated results from the nominal slice thicknesses were within 1.0 mm. The automated results are comparable to those from manual approach (i.e., the difference of both is within 12%). The automatic procedure accurately obtained slice thickness even when the phantom was moved from the iso-center position by up to 4 cm above and 4 cm below the iso-center. The automated results were similar (to within 0.1 mm) for various reconstruction filters. CONCLUSIONS: We successfully developed an automated procedure of slice thickness verification and confirmed that the automated procedure provided accurate results. It provided an easy and effective method of determining slice thickness.


Assuntos
Tomografia Computadorizada por Raios X , Humanos , Imagens de Fantasmas