Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Rev Sci Instrum ; 93(10): 103519, 2022 Oct 01.
Artigo em Inglês | MEDLINE | ID: mdl-36319331

RESUMO

The newly operating near-backscattering imaging (NBI) system on the Laser MegaJoule (LMJ) is briefly described with emphasis on the temporally resolved measurements and their synchronization with the LMJ laser pulse through target shots taken as part of the diagnostic commissioning campaign. The NBI measures the stimulated Brillouin and Raman scattered light around two quadruplets (one inner and one outer) of the upper LMJ hemisphere. The temporal resolution is achieved with a unique system: a specifically designed wide-open optical lens images 40 points of a diffuser onto an array of optical fibers with the scattered light recorded on a multiplexed photodiode array.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...