Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Otolaryngol Head Neck Surg ; 127(6): 520-30, 2002 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-12501103

RESUMO

OBJECTIVE: The purpose of this study was to measure the forces experienced by the SOUNDTEC Direct System magnetic implant during 0.3-T MRI. STUDY DESIGN: Torsional and linear forces imposed on 8 implants were measured by using calibrated neurologic Von Frey hairs and were compared with finite-element analysis predictions and the forces required to separate the incudostapedial joints of unpreserved temporal bones. An implanted embalmed autopsy specimen was also examined before and after 1.5-T MRI. RESULTS: Peak linear force at the orifice of the MRI core measured 0.51g (+/-0.2 SD). Maximum torque occurred at the MRI core center and measured 11.4g-cm (+/-1.2 SD). The mean torque required to separate the incudostapedial joints of 12 unpreserved temporal bones was 33.8g-cm (+/-20.4 SD). The autopsy specimen sustained a 1.5-T MRI scan without disruption of the ossicular chain or explantation. CONCLUSIONS: Physical and mechanical testing of the SOUNDTEC implant indicates that the structural integrity of the ossicles will be maintained during 0.3-T MRI of the human head.


Assuntos
Fenômenos Biomecânicos , Implantes Cocleares/normas , Imageamento por Ressonância Magnética/métodos , Força Compressiva , Qualidade de Produtos para o Consumidor , Humanos , Magnetismo , Desenho de Prótese , Sensibilidade e Especificidade , Osso Temporal/fisiologia
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...