1.
Biol Trace Elem Res ; 43-45: 571-83, 1994.
Artigo em Inglês | MEDLINE | ID: mdl-7710875

RESUMO

The aim of the IRMM--International Measurement Evaluation Program (IMEP) is to test out a possible realization of international measurement comparability for field laboratories through traceability of their measurements to the SI Unit for amount of substance: the mole. In IMEP-3, 10 different trace elements, B, Ca, Cd, Cu, Fe, K, Li, Pb, Rb, and Zn, were determined in a synthetic and a natural water by participating laboratories using their routine methods and graphically compared (in coded form) to certified values, established by IRMM and NIST using an isotope-specific method (Isotope Dilution Mass Spectrometry, Neutron Activation Analysis). The number of participants was 70; 64 laboratories have reported results. The results show a spread of more than 50% asymmetrically around the certified value. The Youden graphs allow evaluation of the overall performance of the laboratories in the IMEP-3 round.


Assuntos
Oligoelementos/análise , Abastecimento de Água/análise , Água/química , Laboratórios/normas , Espectrometria de Massas , Análise de Ativação de Nêutrons , Técnica de Diluição de Radioisótopos , Padrões de Referência