Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Ultrasonics ; 142: 107372, 2024 Jun 04.
Artigo em Inglês | MEDLINE | ID: mdl-38850600

RESUMO

The clinical use of high intensity focused ultrasound (HIFU) therapy for noninvasive tissue ablation has recently gained momentum. Guidance is provided by either magnetic resonance imaging (MRI) or conventional B-mode ultrasound imaging, each with its own advantages and disadvantages. The main limitation of ultrasound imaging is its inability to provide temperature measurements over the ranges corresponding to the target temperatures during ablative thermal therapies (between 55 °C and 70 °C). Here, variations in ultrasound backscattered energy (ΔBSE) were used to monitor temperature increases in liver tissue up to an absolute value of 90 °C during and after HIFU treatment. In vitro experimental measurements were performed in 47 bovine liver samples using a toroidal HIFU transducer operating at 2.5 MHz to increase the temperature of tissues. An ultrasound imaging probe working at 7.5 MHz was placed in the center of the HIFU transducer to monitor the backscattered signals. The free-field acoustic power was set to 9 W, 12 W or 16 W in the different experiments. HIFU sonications were performed for 240 s using a duty cycle of 83 % to allow ultrasound imaging and raw radiofrequency data acquisition during exposures. Measurements showed a linear relationship between ΔBSE (in dB) and temperature (r = 0.94, p < 0.001) over a temperature range from 37 °C to 90 °C, with a high reliability of temperature measurements below 75 °C. Monitoring can be performed at the frame rate of ultrasound imaging scanners with an accuracy within an acceptable threshold of 5 °C, given the temperatures targeted during thermal ablations. If the maximum temperature reached is below 70 °C, ΔBSE is also a reliable approach for estimating the temperature during cooling. Histological analysis shown the impact of the treatment on the spatial arrangement of cells that can explain the observed variation of ΔBSE. These results demonstrate the ability of ΔBSE measurements to estimate temperature in ultrasound images within an effective therapeutic range. This method can be implemented clinically and potentially applied to other thermal-based therapies.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...