1.
Anal Sci ; 22(7): 949-53, 2006 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-16837744

RESUMO

A novel high-selective potentiometric sensor for molybdate was prepared with a PVC membrane combining mu-oxo-bis[5,10,15,20-tetra(p-methylphenyl)porphinatomanganese(III)] [[Mn(p-Me)TPP](2)O] as an electroactive material and 2-nitrophenyl octyl ether (o-NPOE) as a plasticizer in the percentage ratio of 3:65:32, [Mn(p-Me)TPP](2)O:o-NPOE:PVC (w:w). The sensor exhibited a linear response with a Nernstian slope of 30.5 mV per decade within a concentration range of 2.1 x 10(-6) to 1.0 x 10(-1) M MoO4(2-), with a working pH range from 5.0 to 12.5, and a fast response time of less than 15 s. The electrode showed improved selectivity toward molybdate with respect to common coexisting anions compared to monometalloporphyrin counterparts. Several electroactive materials and solvent mediators were compared and the experimental conditions were optimized. The sensor is preliminary applied to the assay of MoO4(2-) in corrosion inhibitor samples with satisfactory results.


Assuntos
Eletrodos Seletivos de Íons , Compostos de Manganês/análise , Manganês/química , Metaloporfirinas/química , Molibdênio/química , Óxidos/análise , Dimerização , Eletrodos , Concentração de Íons de Hidrogênio , Membranas Artificiais , Cloreto de Polivinila/química , Potenciometria/métodos , Solventes/química