Eur Heart J Imaging Methods Pract ; 2(3): qyae097, 2024 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-39391529

RESUMO

Aims: Accurate cardiac chamber quantification is essential for clinical decisions and ideally should be consistent across different echocardiography systems. This study evaluates variations between the Philips EPIQ CVx (version 9.0.3) and Canon Aplio i900 (version 7.0) in measuring cardiac volumes, ventricular function, and valve structures. Methods and results: In this gender-balanced, single-centre study, 40 healthy volunteers (20 females and 20 males) aged 40 years and older (mean age 56.75 ± 11.57 years) were scanned alternately with both systems by the same sonographer using identical settings for both 2D and 4D acquisitions. We compared left ventricular (LV) and right ventricular (RV) volumes using paired t-tests, with significance set at P < 0.05. Correlation and Bland-Altman plots were used for quantities showing significant differences. Two board-certified cardiologists evaluated valve anatomy for each platform. The results showed no significant differences in LV end-systolic volume and LV ejection fraction between platforms. However, LV end-diastolic volume (LVEDV) differed significantly (biplane: P = 0.018; 4D: P = 0.028). Right ventricular (RV) measurements in 4D showed no significant differences, but there were notable disparities in 2D and 4D volumes within each platform (P < 0.01). Significant differences were also found in the LV systolic dyssynchrony index (P = 0.03), LV longitudinal strain (P = 0.04), LV twist (P = 0.004), and LV torsion (P = 0.005). Valve structure assessments varied, with more abnormalities noted on the Philips platform. Conclusion: Although LV and RV volumetric measurements are generally comparable, significant differences in LVEDV, LV strain metrics, and 2D vs. 4D measurements exist. These variations should be considered when using different platforms for patient follow-ups.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA