1.
Ophthalmic Res ; 38(5): 309-12, 2006.
Artigo em Inglês | MEDLINE | ID: mdl-16974133

RESUMO

PURPOSE: To examine the tolerance and mechanical function of an adjustable intraocular lens (IOL) in rabbit eyes. METHODS: Implantation of the *Acri.Tec AR-1 PC/IOL into 14 rabbit eyes. Manipulation of the lens 8 weeks after implantation in order to change the refractive power. Follow-up for up to 5 months. Histopathologic examination of the eyes. RESULTS: Implantation and mechanical adjustment of the PC/IOL were possible. Eyes healed normally. No difference between eyes containing the *Acri.Tec AR-1 PC/IOL and eyes containing the control PC/IOL could be detected with respect to signs of inflammatory reaction, corneal transparency, intraocular pressure and histopathologic appearance. Histopathologic examination of the eyes showed that the *Acri.Tec AR-1 PC/IOL did not cause any damage in rabbit eyes. CONCLUSION: The *Acri.Tec AR-1 PC/IOL is well tolerated in rabbit eyes for extended periods of time, suggesting that this PC/IOL should be well tolerated in the long run. Surgical adjustment of the adjusting element can be performed with little effort several weeks after implantation.