Anaesthesia ; 70(5): 577-84, 2015 May.
Artigo em Inglês | MEDLINE | ID: mdl-25581493

RESUMO

We have used computational fluid dynamic modelling to study the effects of tracheal tube size and position on regional gas flow in the large airways. Using a three-dimensional mathematical model, we simulated flow with and without a tracheal tube, replicating both physiological and artificial breathing. Ventilation through a tracheal tube increased proportional flow to the left lung from 39.5% with no tube to 43.1-47.2%, depending on tube position. Ventilation mode and tube distance from the carina had no effect on flow. Lateral displacement and deflection of the tube increased ventilation to the ipsilateral lung; for example, when deflected 10° to the left of centre, flow to the left lung increased from 43.8 to 53.7%. Because of the small diameter of a tracheal tube relative to the trachea, gas exits a tube at high velocity such that regional ventilation may be affected by changes in the position and angle of the tube.


Assuntos
Manuseio das Vias Aéreas/instrumentação , Intubação Intratraqueal/instrumentação , Respiração Artificial/instrumentação , Manuseio das Vias Aéreas/métodos , Gráficos por Computador , Humanos , Intubação Intratraqueal/métodos , Pulmão/fisiologia , Modelos Anatômicos , Modelos Estatísticos , Respiração Artificial/métodos , Traqueia
SELEÇÃO DE REFERÊNCIAS