1.
Bioengineering (Basel) ; 10(5)2023 Apr 26.
Artigo em Inglês | MEDLINE | ID: mdl-37237593

RESUMO

BACKGROUND: Endoscopic retrograde cholangiopancreatography (ERCP) is crucial to the treatment of biliopancreatic diseases with iatrogenic perforation as a potential complication. As of yet, the wall load during ERCP is unknown, as it is not directly measurable during an ERCP in patients. METHODS: In a life-like, animal-free model, a sensor system consisting of five load cells was attached to the artificial intestines (sensors 1 + 2: pyloric canal-pyloric antrum, sensor 3: duodenal bulb, sensor 4: descending part of the duodenum, sensor 5: distal to the papilla). Measurements were made with five duodenoscopes (n = 4 reusable and n = 1 single use). RESULTS: Fifteen standardized duodenoscopies were performed. Peak stresses were found at the antrum during the gastrointestinal transit (sensor 1 max. 8.95 N, sensor 2 max. 2.79 N). The load reduced from the proximal to the distal duodenum and the greatest load in the duodenum was discovered at the level of the papilla in 80.0% (sensor 3 max. 2.06 N). CONCLUSIONS: For the first time, intraprocedural load measurements and exerting forces obtained during a duodenoscopy for ERCP in an artificial model were recorded. None of the tested duodenoscopes were classified as dangerous for patient safety.