This paper reports an experimental evaluation of the following three different MRI-compatible actuators: a Shinsei ultrasonic motor a Nanomotion ultrasonic motor and a pneumatic cylinder actuator. We report the results of a study comparing the effect of these actuators on the signal to noise ratio (SNR) of MRJ images under a variety of experimental conditions. Evaluation was performed with the controller inside and outside the scanner room and with both 1.5T and 3T MRI scanners. Pneumatic cylinders function with no loss of SNR with controller both inside and outside of the scanner room. The Nanomotion motor performs with moderate loss of SNR when moving during imaging. The Shinsei is unsuitable for motion during imaging. All may be used when motion is appropriately interleaved with imaging cycles.


Assuntos
Artefatos , Análise de Falha de Equipamento , Interpretação de Imagem Assistida por Computador/métodos , Imagem por Ressonância Magnética Intervencionista/instrumentação , Robótica/instrumentação , Cirurgia Assistida por Computador/instrumentação , Desenho de Equipamento , Aumento da Imagem/métodos , Imagem por Ressonância Magnética Intervencionista/métodos , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Cirurgia Assistida por Computador/métodos