1.
Talanta ; 33(6): 537-40, 1986 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-18964135

RESUMO

A method is described for the fluorimetric determination of zinc, based on formation of a zinc-morin complex in the presence of a non-ionic surfactant. The complex has practically no fluorescence in the absence of surfactant, but the addition of Genapol PF-20 (non-ionic surfactant, ethylene oxide-propylene oxide condensate) makes possible the fluorimetric determination of low concentrations of zinc as it enhances the fluorescence of the complex about 75-fold. Maximum fluorescence is produced at pH 4.7 +/- 0.2 (acetic acid-acetate buffer), with 1.5% surfactant and 0.009% morin. The fluorescence is excited at 433 nm and measured at 503 nm. The calibration graph is linear up to 150 ng/ml zinc concentration and the detection limit is 3 ng/ml. The relative standard deviation (11 replicates) is 2.4% for zinc at 20 ng/ml concentration and 1.7% for 100 ng/ml. Of 29 ions studied, Al(3+), Be(2+), Zr(4+) and Cd(2+) strongly increase the fluorescence of the system, and Fe(3+), Ni(2+), Cu(2+), Ti(IV) and Co(2+) decrease the fluorescence signal.