J Innov Card Rhythm Manag
; 14(7): 5499-5503, 2023 Jul.
Artigo
em Inglês
| MEDLINE
| ID: mdl-37492696
RESUMO
Radiofrequency (RF) ablation can be a source of electromagnetic interference (EMI) for cardiovascular implantable electronic devices (CIEDs). The response of CIEDs to this type of EMI can be variable and unpredictable. We report a case with an uncommon response where there was a failure to deliver pacing pulses to both atrial and ventricular pacing leads during RF ablation close to the atrial lead even when the pacemaker was set to pace asynchronously. We also explain the mechanism behind this unusual pacemaker response.