Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Br J Ophthalmol ; 95(4): 498-501, 2011 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-20657018

RESUMO

BACKGROUND: To evaluate the accuracy of an open-field autorefractor compared with subjective refraction in pseudophakes and hence its ability to assess objective eye focus with intraocular lenses (IOLs). METHODS: Objective refraction was measured at 6 m using the Shin-Nippon NVision-K 5001/Grand Seiko WR-5100K open-field autorefractor (five repeats) and by subjective refraction on 141 eyes implanted with a spherical (Softec1 n=53), aspherical (SoftecHD n=37) or accommodating (1CU n=22; Tetraflex n=29) IOL. Autorefraction was repeated 2 months later. RESULTS: The autorefractor prescription was similar (average difference: 0.09 ± 0.53 D; p=0.19) to that found by subjective refraction, with ∼71% within ± 0.50 D. The horizontal cylindrical components were similar (difference: 0.00 ± 0.39 D; p=0.96), although the oblique (J(45)) autorefractor cylindrical vector was slightly more negative (by -0.06 ± 0.25 D; p=0.06) than the subjective refraction. The results were similar for each of the IOL designs except for the spherical IOL, where the mean spherical equivalent difference between autorefraction and subjective was more hypermetropic than the Tetraflex accommodating IOL (F=2.77, p=0.04). The intrasession repeatability was <0.55 D (95% CI) and intersession repeatability <0.50 D in ≥ 85%. CONCLUSIONS: The autorefractor gives valid and repeatable measures of pseudophakic eye refraction and hence objective accommodation.


Assuntos
Acomodação Ocular/fisiologia , Dispositivos Ópticos/normas , Pseudofacia/fisiopatologia , Refração Ocular/fisiologia , Erros de Refração/diagnóstico , Seleção Visual/instrumentação , Idoso , Feminino , Humanos , Lentes Intraoculares , Masculino , Erros de Refração/fisiopatologia , Reprodutibilidade dos Testes , Seleção Visual/normas , Acuidade Visual/fisiologia
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...