Ophthalmic Physiol Opt ; 28(1): 29-34, 2008 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-18201333

RESUMO

BACKGROUND: The purpose of this study was to assess the accuracy of the ICare tonometer, using the Perkins applanation tonometer (AT) as the reference, in a sample population being treated with travaprost 0.004% for glaucoma. MATERIAL AND METHODS: Twenty-eight consecutive patients with open-angle glaucoma or ocular hypertension who had been receiving travaprost 0.004% to control elevated intraocular pressure (IOP) were included in the study. IOP was measured in the entire sample with ICare and Perkins AT. The difference between the methods was plotted against the mean to compare the tonometers. The hypothesis of zero bias was examined by a paired t-test. The 95% limits of agreement (LoA) were also calculated. RESULTS: As previously found in young healthy subjects, ICare showed a tendency to overestimate Perkins IOP measurements by a mean of 3.57 mmHg. The agreement between the two methods is shown by the 95% LoA which was from -9.37 to +2.23 mmHg: 53% of the IOP differences fell within +/-3 mmHg. CONCLUSIONS: The performance of the ICare tonometer in glaucomatous patients being treated with travaprost 0.004% to lower the IOP appears to be similar to that of young healthy patients. The tendency of ICare to overestimate the IOP readings should be considered when the instrument is used in the follow-up of glaucomatous patients.


Assuntos
Anti-Hipertensivos/uso terapêutico , Cloprostenol/análogos & derivados , Pressão Intraocular/efeitos dos fármacos , Hipertensão Ocular/diagnóstico , Tonometria Ocular/instrumentação , Adulto , Idoso , Idoso de 80 Anos ou mais , Cloprostenol/uso terapêutico , Desenho de Equipamento , Feminino , Glaucoma de Ângulo Aberto/diagnóstico , Glaucoma de Ângulo Aberto/tratamento farmacológico , Humanos , Masculino , Pessoa de Meia-Idade , Hipertensão Ocular/tratamento farmacológico , Valores de Referência , Travoprost
2.
Optom Vis Sci ; 83(2): 102-7, 2006 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-16501412

RESUMO

PURPOSE: The ICare (Tiolat Oy, Helsinki, Finland) is a new portable tonometer that measures intraocular pressure (IOP) with a new rebound method, in which a very light probe is used to make momentary contact with the cornea in slow motion. The purpose of this study is to assess the accuracy of the ICare IOP measurements by comparing them against other portable tonometers: Perkins applanation tonometer and Tono-Pen XL digital tonometer (Medtronic Solan, Jacksonville, FL). METHODS: Sixty-five young subjects were assessed with each of the tonometers. ICare tonometry was performed first, followed by Perkins applanation tonometry and Tono-Pen XL in a random order. Regression analysis was used to evaluate the relationship between the Perkins tonometer and the remaining tonometers used in this study. Tonometers were also compared by plotting the difference between the methods against the mean. The hypothesis of zero bias was examined by a paired t test and the 95% limits of agreement (LoA) were calculated. RESULTS: ICare and Tono-Pen XL significantly overestimate IOP when compared with Perkins applanation tonometry. The mean of the difference between Perkins and ICare and Perkins and Tono-Pen XL was (mean +/- standard deviation) -3.35 +/- 2.28 mm Hg and -2.78 +/- 2.53 mm Hg, respectively. The 95% LoA between Perkins tonometry and ICare tonometry were between -7.81 and +1.12 and between Perkins tonometry and Tono-Pen XL tonometry between -7.74 and +2.18. CONCLUSIONS: Compared with Perkins tonometry, the ICare tonometer allows clinicians to estimate IOP with a portable, rapid, and noninvasive method with similar reliability to that offered by Tono-Pen XL. Clinicians should be aware of the systematic overestimation of IOP with the ICare. Further research is needed to evaluate the performance of rebound tonometry in populations with higher IOP and assess the reliability of this technique in the early detection and follow up of glaucomatous patients.


Assuntos
Pressão Intraocular/fisiologia , Tonometria Ocular , Adulto , Desenho de Equipamento , Feminino , Humanos , Masculino , Valores de Referência , Reprodutibilidade dos Testes , Tonometria Ocular/instrumentação , Tonometria Ocular/normas