RESUMO

Liquid biopsies are an integral part of the diagnosis of cancer. Here, we have extended previous validation studies of a new targeted NGS panel to include the detection of copy number amplifications (CNAs), fusions, and exon skipping variants. Detection of these gene classes included specimens from clinical and healthy donors and cell lines (fusions: ROS1, EML4-ALK, NTRK1; exon skipping: MET exon 14; CNAs: HER2, CDK6, EGFR, MYC, and MET). The limit of detection (LOD) for fusion/skipping was 42 copies (QC threshold was three copies) and was verified using three additional fusion/skipping variants. LOD for CNAs was 1.40-fold-change (QC threshold = 1.15-fold change) and was verified with three additional CNAs. In repeatability and intermediate precision (within lab) studies, all fusion/skipping variants were detected in all runs and all days of testing (n = 18/18; 100%); average CV for repeatability was 20.5% (range 8.7-34.8%), and for intermediate precision it was 20.8% (range 15.7-30.5%). For CNAs, 28/29 (96.6%) copy gains were detected. For CNAs, the average CV was 1.85% (range 0% to 5.49%) for repeatability and 6.59% (range 1.65% to 9.22%) for intermediate precision. The test panel meets the criteria for being highly sensitive and specific and extends its utility for the serial detection of clinically relevant variants in cancer.