Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 10 de 10
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Artif Organs ; 27(11): 1050-3, 2003 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-14616524

RESUMO

Silicone rubber hollow fiber membrane produces an ideal gas exchange for long-term ECMO due to nonporous characteristics. The extracapillary type silicone rubber ECMO oxygenator having an ultrathin hollow fiber membrane was developed for pediatric application. The test modules were compared to conventional silicone coil-type ECMO modules. In vitro experiments demonstrated a higher O2 and CO2 transfer rate, lower blood flow resistance, and less hemolysis than the conventional silicone coil-type modules. This oxygenator was combined with the Gyro C1E3 centrifugal pump, and three ex vivo experiments were conducted to simulate pediatric V-A ECMO condition. Four day and 6 day experiments were conducted in cases 1 and 2, respectively. Case 3 was a long-term experiment up to 2 weeks. No plasma leakage and stable gas performances were achieved. The plasma free hemoglobin was maintained within a normal range. This compact pump-oxygenator system in conjunction with the Gyro C1E3 centrifugal pump has potential for a hybrid total ECMO system.


Assuntos
Oxigenação por Membrana Extracorpórea , Oxigenadores de Membrana , Animais , Bovinos , Hemoglobinas/análise , Teste de Materiais , Elastômeros de Silicone
2.
ASAIO J ; 49(1): 112-6, 2003.
Artigo em Inglês | MEDLINE | ID: mdl-12558317

RESUMO

Centrifugal blood pumps are used widely for cardiopulmonary bypass, as ventricular assist devices, and for extracorporeal membrane oxygenation (ECMO). However, there is no centrifugal blood pump that is suitable for long-term ECMO. The authors developed the Gyro C1E3 centrifugal blood pump (Kyocera Corporation, Kyoto, Japan), which has superior antithrombogenic, antitraumatic, and hydraulic features in comparison with the conventional centrifugal blood pumps. Five ex vivo long-term durability tests of the Gyro C1E3 were performed using healthy miniature calves. The ECMO circuit was composed of a prototype hollow fiber silicone membrane oxygenator and a Gyro C1E3 pump. Venous blood was drained from the left jugular vein of a calf, passed through the oxygenator and infused into the left carotid artery using a Gyro C1E3. Ex vivo studies were performed from 7 to 15 days at a blood flow rate of 1 L/min. During this period, the Gyro C1E3 demonstrated a stable performance without exchanging the pump. Bleeding complications were the major reason for termination of each experiment. Rotational speed was maintained around 2,000 rpm. All five calves demonstrated neither abnormal signs nor abnormal blood examination data throughout the experiment. Neither clot nor thrombus formations were found during the necropsy in the cannula or pump nor were infarctions observed in any of the major organs. In conclusion, the Gyro C1E3 showed a stable and reliable performance during long-term ex vivo bovine experiments under the conditions tested.


Assuntos
Ponte Cardiopulmonar/instrumentação , Coração Auxiliar , Animais , Bovinos , Centrifugação , Desenho de Equipamento , Contagem de Eritrócitos , Contagem de Leucócitos
3.
Ann Thorac Cardiovasc Surg ; 8(5): 268-74, 2002 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-12472408

RESUMO

Throughout the last 50 years, many improvements have been made for a more effective oxygenator. A large plate type membrane oxygenator, used by Clowes, and a coil type, used by Kolff, has evolved into the small hollow fiber oxygenator. The complex bubble oxygenator, or rotating disk oxygenator, has become a small disposable bubble oxygenator. The currently available oxygenators are extremely small, efficient, and can be used for extended periods of time. However, there are some problems with extracorporeal membrane oxygenation (ECMO). Currently in the United States, there are no clinically applicable hollow fiber ECMO oxygenators available, in spite of the extended ECMO application. Therefore, the development of a small, yet efficient, silicone hollow fiber membrane oxygenator for long-term ECMO usage was attempted. Based on the results of many experimental models, preclinical oxygenator models for long-term ECMO were developed in our laboratory using an ultra-thin silicone rubber hollow fiber membrane.


Assuntos
Oxigenação por Membrana Extracorpórea/instrumentação , Oxigenadores de Membrana , Desenho de Equipamento , Teste de Materiais , Silicones
4.
ASAIO J ; 48(6): 636-9, 2002.
Artigo em Inglês | MEDLINE | ID: mdl-12455774

RESUMO

A connection was previously reported between the hemolytic characteristics associated with oxygenators and the pressure drop measurements in the blood chamber under experimental conditions simulating their use in cardiopulmonary bypass. We examined this association during extracorporeal membrane oxygenation (ECMO) conditions. Three oxygenators for ECMO or pediatric cardiopulmonary bypass (Menox EL4000, Dideco Module 4000, and Mera HPO-15H) were evaluated. Fresh blood from healthy Dexter strain calves anticoagulated with citrate phosphate dextrose adenine solution was used. The blood flow was fixed at 1 L/min, similar to that in ECMO. The Normalized Index of Hemolysis for Oxygenators (NIHO) has been modified according to the American Society of Testing and Materials standards, as was previously reported. The NIHO value was the lowest in the Menox (0.0070+/-0.0009) and increased from Menox to Dideco (0.0113+/-0.0099) to Mera (0.0164+/-0.0043); however, there were no significant differences among the oxygenators. This NIHO value has a close correlation to the pressure drop. In conclusion, this evaluation method is also applicable to comparison of the biocompatibility performance of different types of clinically available oxygenators for ECMO.


Assuntos
Oxigenação por Membrana Extracorpórea/métodos , Hemólise , Animais , Velocidade do Fluxo Sanguíneo , Pressão Sanguínea , Ponte Cardiopulmonar/instrumentação , Bovinos , Oxigenação por Membrana Extracorpórea/normas , Valores de Referência
5.
J Med Invest ; 49(3-4): 156-62, 2002 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-12323005

RESUMO

Currently in United States, there are no clinically-applicable hollow fiber extracorporeal membrane oxygenation (ECMO) oxygenators available. Therefore, our laboratory is in the process of developing a silicone hollow fiber membrane oxygenator for long-term ECMO usage. This oxygenator incorporates an ultrathin silicone hollow fiber. At this time, a specially-modified blood flow distributor (one chamber distributor) is centered in the module to prevent blood stagnation. An ex vivo long-term durability test for ECMO was performed using a healthy miniature calf for 2 weeks. Venous blood was drained from the left jugular vein of a calf, passed through the oxygenator and infused into the left carotid artery using a Gyro C1E3 centrifugal blood pump. A successful 2-week ex vivo experiment was performed. The O2 and CO2 gas transfer rates were maintained at the same value of 40 m/min at a blood flow rate of 1 L/min flow and V/Q=3 (V=gas flow rate; Q=blood flow rate). The plasma free hemoglobin was maintained around 5 mg/dl. After the experiment, no blood clot formation was observed in the module and no abnormal necropsy findings were found. These data suggest that the performance of this newly-improved oxygenator was stable, reliable, and acceptable for long-term ECMO.


Assuntos
Oxigenação por Membrana Extracorpórea/instrumentação , Animais , Contagem de Células Sanguíneas , Proteínas Sanguíneas/análise , Dióxido de Carbono/sangue , Bovinos , Desenho de Equipamento , Estudos de Viabilidade , Feminino , Oxigênio/sangue , Pressão , Silicones
6.
Artif Organs ; 26(1): 45-8, 2002 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-11872011

RESUMO

We are developing the Baylor-Kyocera KP implantable centrifugal blood pump for small sized adult and pediatric patients. This pump eccentrically positions the inlet port, which eliminates flow stagnation around the top pivot bearing. The inlet port design is important because it may vary the inlet orifice pressure on the top housing and change hydraulic performance and hemolytic characteristics. The pressure distribution inside the KP pump was assessed by a computational fluid dynamic (CFD) analysis with 2.7 x 10(5) elements and 3.16 x 10(5) nodes. Hydraulic performance and hemolysis were evaluated with 3 different pump housings, which had 3.8, 4.5, and 6.1 mm offset inlet ports from the center in a mock circuit. The CFD analysis revealed that the pressure gradually increased from the center toward the peripheral. The pressure difference between the 3.8 to 6.1 mm offsets was less than 600 Pa. The hydraulic performance did not drastically change at 3.8, 4.5, and 6.1 mm offset from the center. However, the hemolysis increased with the increase of the port offset from 0.0080+/- 0.0048 to 0.054 +/- 0.028 g/100 L. The inlet port positioning is important to attain less blood trauma in this small Gyro centrifugal blood pump. The preferable position of the inlet port is less than 4.5 mm offset from the center.


Assuntos
Coração Auxiliar , Centrifugação , Desenho de Equipamento , Humanos
7.
Artif Organs ; 26(1): 67-71, 2002 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-11872016

RESUMO

To design a miniaturized centrifugal blood pump, the impeller internal diameter (ID), which is a circle diameter on the inner edge of the vane, is considered one of the important aspects. Hydraulic performance, hemolysis, and thrombogenicity were evaluated with different impeller IDs. Two impellers were fabricated with an outer diameter of 35 mm, of which 1 had an 8 mm ID impeller and the other had a 12 mm ID. These impellers were combined with 2 different housings in which the inlet port was eccentrically positioned 3.8 and 4.5 mm offset from the center. The hydraulic performance and hemolysis were evaluated in a mock circuit, and thrombogenicity was evaluated in a 2 day ex vivo study with each impeller housing combination. Both impellers required 3,000 rpm in the 3.8 mm offset inlet to attain 5 L/min against 100 mm Hg (left ventricular assist device condition). The 8 mm ID impeller required 3,200 rpm, and the 12 mm ID impeller required 3,100 rpm in the 4.5 mm offset housing. The normalized index of hemolysis was 0.0080 +/- 0.0048 g/100 L in the 8 mm ID impeller with the 3.8 mm offset and 0.022 +/- 0.018 g/100 L with 4.5 mm offset. The 12 mm ID impeller had 0.068 +/- 0.028 g/100 L with the 3.8 mm offset and 0.010 +/- 0.002 g/100 L with the 4.5 mm offset. After the 2 day ex vivo study, no blood clot was formed around the top bearing in all the pump heads. The 8 mm ID impeller with 3.8 mm offset inlet and the 12 mm ID impeller with the 4.5 mm offset had less hemolysis compared to the other pump heads that were subjected to 14 day ex vivo and 10 day in vivo studies. The 8 mm ID impeller with the 3.8 mm offset inlet had a blood clot around the top bearing after the 14 day ex vivo study. No thrombus was found around the top bearing of the 12 mm ID impeller with the 4.5 mm offset in the 10 day in vivo study. These results suggest that the ID does not greatly change the hydraulic performance of a small centrifugal blood pump. The proper combination of the impeller ID and inlet port offset obtains less hemolysis. The larger impeller ID is considered to have less thrombogenicity around the top bearing.


Assuntos
Coração Auxiliar , Centrifugação , Hemólise , Humanos , Miniaturização , Rotação
8.
Ann Thorac Cardiovasc Surg ; 8(1): 7-11, 2002 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-11916436

RESUMO

Based on the results of many experimental models, a hollow fiber silicone membrane oxygenator applicable for long-term extracorporeal membrane oxygenation (ECMO) was developed. For further high performance and antithrombogenicity, this preclinical model was modified, and a new improved oxygenator was successfully developed. In addition to ECMO application, the superior biocompatibility of silicone must be advantageous for pediatric cardiopulmonary bypass (CPB). An ex vivo short-term durability test for pediatric CPB was performed using a healthy miniature calf for six hours. Venous blood was drained from the left jugular vein of a calf, passed through the oxygenator and infused into the left carotid artery using a Gyro C1E3 centrifugal pump. For six hours, the O2 and CO2 gas transfer rates were maintained around 90 and 80 ml/min at a blood flow rate of 2 L/min and V/Q=3, respectively. The plasma free hemoglobin was maintained around 5 mg/dl. These data suggest that this newly improved oxygenator has superior efficiency, less blood trauma, and may be suitable for not only long-term ECMO but also pediatric CPB usage.


Assuntos
Ponte Cardiopulmonar/instrumentação , Oxigenação por Membrana Extracorpórea/instrumentação , Oxigenadores de Membrana , Animais , Bovinos , Desenho de Equipamento , Pediatria/instrumentação , Silicones
9.
Artif Organs ; 20(5): 485-490, 1996 May.
Artigo em Inglês | MEDLINE | ID: mdl-28868689

RESUMO

Since 1991, in our laboratory, a pivot bearing-supported, sealless, centrifugal pump has been developed as an implantable ventricular assist device (VAD). For this application, the configuration of the total pump system should be relatively small. The C1E3 pump developed for this purpose was anatomically compatible with the small-sized patient population. To evaluate an-tithrombogenicity, ex vivo 2-week screening studies were conducted instead of studies involving an intracorpore-ally implanted VADs using calves. Five paracorporeal LVAD studies were performed using calves for longer than 2 weeks. The activated clotting time (ACT) was maintained at approximately 250 s using heparin. All of the devices demonstrated trouble-free performances over 2 weeks. Among these 5 studies, 3 implantations were subjected to 1-month system validation studies. There were no device-induced thrombus formations inside the pump housing, and plasma-free hemoglobin levels in calves were within the normal range throughout the experiment (35, 34, and 31 days). There were no incidents of system malfunction. Subsequently, the mass production model was fabricated and yielded a normalized index of hemolysis of 0.0014, which was comparable to that of clinically available pumps. The wear life of the impeller bearings was estimated at longer than 8 years. In the next series of in vivo studies, an implantable model of the C1E3 pump will be fabricated for longer term implantation. The pump-actuator will be implanted inside the body; thus the design calls for substituting plastic for metallic parts.

10.
Artif Organs ; 20(5): 605-612, 1996 May.
Artigo em Inglês | MEDLINE | ID: mdl-28868707

RESUMO

Two well-known centrifugal and axial pumping principles are used simultaneously in a new blood pump design. Inside the pump housing is a spiral impeller, a conically shaped structure with threads on the surface. The worm gears provide an axial motion of the blood column through the threads of the central cone. The rotational motion of the conical shape generates the centrifugal pumping effect and improves the efficiency of the pump without increasing hemolysis. The hydrodynamic performance of the pump was examined with a 40% glycerin-water solution at several rotation speeds. The gap between the housing and the top of the thread is a very important factor: when the gap increases, the hydrodynamic performance decreases. To determine the optimum gap, several in vitro hemolysis tests were performed with different gaps using bovine blood in a closed circuit loop under two conditions. The first simulated condition was a left ventricular assist device (LVAD) with a flow rate of 5 L/min against a pressure head of 100 mm Hg, and the second was a cardiopulmonary bypass (CPB) simulation with a flow rate of 5 L/min against 350 mm Hg of pressure. The best hemolysis results were seen at a gap of 1.5 mm with the normalized index of hemolysis (NIH) of 0.0063 ± 0.0020 g/100 L and 0.0251 ± 0.0124 g/100 L (mean ± SD; n = 4) for LVAD and CPB conditions, respectively.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...