1.
Artigo em Inglês | MEDLINE | ID: mdl-38907647

RESUMO

Miniscrews are temporary skeletal anchorage devices that are widely used in orthodontic treatment, and their success depends on the placement area, angle, technique, and screw dimensions. This study aimed to investigate the effects of miniscrew lengths, insertion angles, and force directions on a mandible model consisting of teeth, cortical and cancellous bones. One Dental Volumetric Tomography (DVT) scan from a patient who had miniscrews were used for mandibular bone modeling to perform finite element analysis. The model variables included miniscrew lengths (6, 8, and 10 mm), insertion angles (-15°, 45°, 60°, and 90°), and force directions (30°, 45°, and 60°). The minimum and maximum stresses were calculated as 18.61 and 37.11 MPa at 6 mm and 10 mm, respectively. According to the insertion angles, the lowest stress was observed at 60°, while the highest stress was found at 15° in the ventral direction. At force directions, the lowest stress was at 60°, and the highest stress was at 45°. However, there were no significant differences in insertion angles and force directions. A statistically significant difference was determined in miniscrew length. As a result, the best result was calculated to be 6 mm inserted at a 60° angle, which could induce the lowest stress. Increasing the miniscrew length will increase the stress on the mandible. In addition, because of the higher force direction, stress decreases with shorter power arms.