Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Intensive Care Med Exp ; 2(1): 10, 2014 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-26266911

RESUMO

BACKGROUND: End-expiratory lung volume measurement by the nitrogen washin-washout technique (EELVWI-WO) may help titrating positive end-expiratory pressure (PEEP) during acute respiratory distress syndrome (ARDS). Validation of this technique has been previously performed using computed tomography (EELVCT), but at mild PEEP levels, and relatively low fraction of inspired oxygen (FiO2), which may have insufficiently challenged the validity of this technique. The aims of this study were (1) to evaluate the reliability of EELVWI-WO measurements at different PEEP and V T during experimental ARDS and (2) to evaluate trending ability of EELVWI-WO to detect EELV changes over time. METHODS: ARDS was induced in 14 piglets by saline lavage. Optimal PEEP was selected during a decremental PEEP trial, based on best compliance, best EELVWI-WO, or a PEEP-FiO2 table. Eight V T (4 to 20 mL · kg(-1)) were finally applied at optimal PEEP. EELVWI-WO and EELVCT were determined after ARDS onset, at variable PEEP and V T. RESULTS: EELVWI-WO underestimated EELVCT with a non-constant linear bias, as it decreased with increasing EELV. Limits of agreement for bias were ±398 mL. Bias between methods was greater at high PEEP, and further increased when high PEEP was combined with low V T. Concordance rate of EELV changes between consecutive measurements was fair (79%). Diagnostic accuracy was good for detection of absolute EELV changes above 200 mL (AUC = 0.79). CONCLUSIONS: The reliability of the WI-WO technique is critically dependent on ventilatory settings, but sufficient to accurately detect EELV change greater than 200 mL.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...