1.
Anaesthesia ; 59(4): 390-3, 2004 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-15023111

RESUMO

Continuous peripheral oxygen saturation monitoring using a finger pulse oximeter is standard in prehospital emergency medicine. Forehead peripheral oxygen saturation monitoring has been enhanced for better performance during movement and in cold ambient temperatures, both of which are common during emergency transport. We compared a new forehead monitor with standard finger pulse oximeter. The forehead technique had significantly fewer mean (SD) alarms per patient (3.0 (2.2)) than the finger pulse oximeter (7.8 (4.0)) and shorter durations of malfunction (76 (60) s compared to 333 (170) s) when using the finger pulse oximeter. We conclude that measuring peripheral oxygen saturation monitoring with a forehead sensor provides better monitoring quality in emergency care.


Assuntos
Dedos/irrigação sanguínea , Testa/irrigação sanguínea , Oxigênio/sangue , Transporte de Pacientes , Adulto , Idoso , Idoso de 80 Anos ou mais , Antropometria , Monitorização Transcutânea dos Gases Sanguíneos/instrumentação , Monitorização Transcutânea dos Gases Sanguíneos/métodos , Emergências , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Pressão Parcial
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...