RESUMO
We present a description of the capabilities and performance of the NAval Ultra-Trace Isotope Laboratory's Universal Spectrometer (NAUTILUS) at the U.S. Naval Research Laboratory. The NAUTILUS combines secondary ion mass spectrometry (SIMS) and single-stage accelerator mass spectrometry (SSAMS) into a single unified instrument for spatially resolved trace element and isotope analysis. The NAUTILUS instrument is essentially a fully functional SIMS instrument with an additional molecule-filtering detector, the SSAMS. The combination of these two techniques mitigates the drawbacks of each and enables new measurement paradigms for SIMS-like microanalysis. Highlighted capabilities include molecule-free raster ion imaging for direct spatially resolved analysis of heterogeneous materials with or without perturbed isotopic compositions. The NAUTILUS' sensitivity to trace elements is at least 10× better than commercial SIMS instruments due to near-zero background conditions. We describe the design and construction of the NAUTILUS, and its performance applied to topics in nuclear materials analysis, cosmochemistry, and geochemistry.
RESUMO
A commercial secondary ion mass spectrometer (SIMS) was coupled to a ± 300 kV single-stage accelerator mass spectrometer (SSAMS). Positive secondary ions generated with the SIMS were injected into the SSAMS for analysis. This combined instrument was used to measure the uranium isotopic ratios in particles of three certified reference materials (CRM) of uranium, CRM U030a, CRM U500, and CRM U850. The ability to inject positive ions into the SSAMS is unique for AMS systems and allows for simple analysis of nearly the entire periodic table because most elements will readily produce positive ions. Isotopic ratios were measured on samples of a few picograms to nanograms of total U. Destruction of UH(+) ions in the stripper tube of the SSAMS reduced hydride levels by a factor of â¼3 × 10(4) giving the UH(+)/U(+) ratio at the SSAMS detector of â¼1.4 × 10(-8). These hydride ion levels would allow the measurement of (239)Pu at the 10 ppb level in the presence of U and the equivalent of â¼10(-10 236)U concentration in natural uranium. SIMS-SSAMS analysis of solid nuclear materials, such as these, with signals nearly free of molecular interferences, could have a significant future impact on the way some measurements are made for nuclear nonproliferation.