Sensors (Basel) ; 23(12)2023 Jun 14.
Artigo em Inglês | MEDLINE | ID: mdl-37420725

RESUMO

The ActiGraph GT9X has been implemented in clinical trials to track physical activity and sleep. Given recent incidental findings from our laboratory, the overall aim of this study was to notify academic and clinical researchers of the idle sleep mode (ISM) and inertial measurement unit (IMU)'s interaction, as well as their subsequent effect on data acquisition. Investigations were undertaken using a hexapod robot to test the X, Y and Z sensing axes of the accelerometers. Seven GT9X were tested at frequencies ranging from 0.5 to 2 Hz. Testing was performed for three sets of setting parameters: Setting Parameter 1 (ISMONIMUON), Setting Parameter 2 (ISMOFFIMUON), Setting Parameter 3 (ISMONIMUOFF). The minimum, maximum and range of outputs were compared between the settings and frequencies. Findings indicated that Setting Parameters 1 and 2 were not significantly different, but both were significantly different from Setting Parameter 3. Upon inspection, it was discovered that the ISM was only active during Setting Parameter 3 testing, despite it being enabled in Setting Parameter 1. Researchers should be aware of this when conducting future research using the GT9X.


Assuntos
Acelerometria , Exercício Físico , Sono
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...