Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 5 de 5
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
PLoS One ; 16(9): e0256514, 2021.
Artigo em Inglês | MEDLINE | ID: mdl-34587170

RESUMO

PURPOSE: Limited work has been performed for the implementation of digital breast tomosynthesis (DBT) in breast cancer surveillance imaging. The aim of this study was to investigate the differences between two different DBT implementations in breast cancer surveillance imaging, for patients with a personal history of breast cancer. METHOD: The DBT implementations investigated were: (1) 2-view 2D digital mammography and 2-view DBT (2vDM&2vDBT) (2) 1-view (cranial-caudal) DM and 1-view (mediolateral-oblique) DBT (1vDM&1vDBT). Clinical performance of these two implementations was assessed retrospectively using observer studies with 118 sets of real patient images, from a single imaging centre, and six observers. Sensitivity, specificity and area under the curve (AUC) using the Jack-knife alternative free-response receiver operating characteristics (JAFROC) analysis were evaluated. RESULTS: Results suggest that the two DBT implementations are not significantly different in terms of sensitivity, specificity and AUC. When looking at the two main different lesion types, non-calcifications and calcifications, and two different density levels, no difference in the performance of the two DBT implementations was found. CONCLUSIONS: Since 1vDM&1vDBT exposes the patient to half the dose of 2vDM&2vDBT, it might be worth considering 1vDM&1vDBT in breast cancer surveillance imaging. However, larger studies are required to conclude on this matter.


Assuntos
Neoplasias da Mama/diagnóstico , Mama/diagnóstico por imagem , Imageamento Tridimensional/métodos , Mamografia/métodos , Adulto , Idoso , Idoso de 80 Anos ou mais , Mama/patologia , Neoplasias da Mama/diagnóstico por imagem , Neoplasias da Mama/patologia , Feminino , Humanos , Pessoa de Meia-Idade , Intensificação de Imagem Radiográfica
2.
Cancer Manag Res ; 12: 9953-9964, 2020.
Artigo em Inglês | MEDLINE | ID: mdl-33116852

RESUMO

PURPOSE: Breast cancer brain metastasis (BCBM) represents a major clinical challenge. Can MRI help in advancements in the management of BCBM? This review discusses MRI developments and the corresponding potential advancements in BCBM management. METHODS: An exhaustive literature search was undertaken to identify studies which look into the potential of MRI in BCBM management. Seven hundred and eighty-four studies published from September 1984 to May 2020 were identified. Three topics are covered where MRI is not clinically established yet: 1) the prognosis of BCBM; 2) the screening of BC patients for BCBM development, and 3) the assessment of imaging features correlated to BC subtype. RESULTS: Thirty-six studies were considered eligible for the purposes of this review. On-going progress is made with the identification of different BCBM characteristics and MRI metrics that might be related to prognosis. Progress has been made with the identification of different BCBM characteristics, including BCBM location, degree of edema, white matter disruption, tumor edge sharpness, and temporal muscle thickness. A more accurate prediction of prognosis could lead to more suitable patient management and treatment. The use of MRI in BCBM screening of the high-risk breast cancer population remains a controversial subject. To date, there are no results from clinical trials; however, there is a rising number of relatively small studies that show concern on this subject and support BCBM screening. It is important to oncologists to be able to assess the tumor subtype non-invasively. MRI features, which have shown some correlation with subtype, include the number of tumors, location, and their distribution in the brain. Advanced tools and metrics have been produced to carry out radiological characteristics analysis on MRI images. Assessing MRI features in more detail could provide a more personalized management of patients. CONCLUSION: Developments in the use of MRI have the potential to improve BCBM management.

3.
Phys Med ; 57: 25-32, 2019 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-30738528

RESUMO

Digital breast tomosynthesis (DBT) is currently under consideration for replacement of, or combined use with 2D-mammography in national breast screening programmes. To investigate the potential benefits that DBT can bring to screening, the threshold detectable lesion diameters were measured for different forms of DBT in comparison to 2D-mammography. The aim of this study was to compare the threshold detectable mass diameters obtained with narrow angle (15°/15 projections) and wide angle (50°/25 projections) DBT in comparison to 2D-mammography. Simulated images of 60 mm thick compressed breasts were produced with and without masses using a set of validated image modelling tools for 2D-mammography and DBT. Image processing and reconstruction were performed using commercial software. A series of 4-alternative forced choice (4AFC) experiments was conducted for signal detection with the masses as targets. The threshold detectable mass diameter was found for each imaging modality with a mean glandular dose of 2.5 mGy. The resulting values of the threshold diameter for 2D-mammography (10.2 ±â€¯1.4 mm) were found to be larger (p < 0.001) than those for narrow angle DBT (6.0 ±â€¯1.1 mm) and wide angle DBT (5.6 ±â€¯1.2 mm). There was no significant difference between the threshold diameters for wide and narrow angle DBT. Implications for the introduction of DBT alone or in combination with 2D-mammography in breast cancer screening are discussed.


Assuntos
Neoplasias da Mama/diagnóstico por imagem , Neoplasias da Mama/patologia , Mamografia/métodos , Carga Tumoral , Humanos
4.
Phys Med Biol ; 62(6): 2376-2397, 2017 03 21.
Artigo em Inglês | MEDLINE | ID: mdl-28151431

RESUMO

In-depth evaluation of the noise and sharpness characteristics of FujiFilm Innovality, GE SenoClaire, Hologic Selenia Dimensions and Siemens Inspiration digital breast tomosynthesis (DBT) systems was performed with the intention of improving image simulation for virtual clinical trials. Noise power spectra (NPS) and modulation transfer function curves (MTF) were measured for planar modes and for the first and central projections for DBT modes. In DBT mode, the x-ray beam was blocked for the projections before the central projection in order to remove the influence of lag and ghosting from the previous images. A quadratic fit between the NPS and linearised pixel value gave the noise coefficients for planar and DBT imaging modes. The spatial frequencies corresponding to an MTF of 0.5 (MTF0.5) were calculated from the MTF measurements made on the breast support and at 40 mm above the breast support. This was done for the first and the central projections. The percentage of signal carried over from the first projection to subsequent images (lag) was measured using a slit. The noise associated with lag was also evaluated. The DBT modes typically had lower electronic noise coefficients but higher structural noise coefficients compared to the respective planar mode MTF0.5 measured 40 mm above the table was between 6% and 47% lower for continuous scanning systems compared to 1% lower for step and shoot systems. For wide angle DBT, the MTF0.5 of the first projection was 18% (FujiFilm) and 28% (Siemens) lower than for the central projection. Lag in the second projection was 2.2%, 0.3%, 0.8% for the FujiFilm, GE and Hologic systems respectively. In all cases, the noise associated with lag was negligible. Current modelling frameworks for virtual clinical trials of breast DBT systems need to be adapted to account for signals from lag and variations in the MTF at wide angles.


Assuntos
Mamografia/métodos , Intensificação de Imagem Radiográfica/métodos , Tomografia por Raios X/métodos , Ensaios Clínicos como Assunto , Feminino , Humanos , Mamografia/normas , Imagens de Fantasmas , Intensificação de Imagem Radiográfica/normas , Razão Sinal-Ruído , Tomografia por Raios X/normas
5.
Phys Med Biol ; 62(3): 858-877, 2017 02 07.
Artigo em Inglês | MEDLINE | ID: mdl-28072582

RESUMO

Digital breast tomosynthesis (DBT) is under consideration to replace or to be used in combination with 2D-mammography in breast screening. The aim of this study was the comparison of the detection of microcalcification clusters by human observers in simulated breast images using 2D-mammography, narrow angle (15°/15 projections) and wide angle (50°/25 projections) DBT. The effects of the cluster height in the breast and the dose to the breast on calcification detection were also tested. Simulated images of 6 cm thick compressed breasts were produced with and without microcalcification clusters inserted, using a set of image modelling tools for 2D-mammography and DBT. Image processing and reconstruction were performed using commercial software. A series of 4-alternative forced choice (4AFC) experiments was conducted for signal detection with the microcalcification clusters as targets. Threshold detectable calcification diameter was found for each imaging modality with standard dose: 2D-mammography: 2D-mammography (165 ± 9 µm), narrow angle DBT (211 ± 11 µm) and wide angle DBT (257 ± 14 µm). Statistically significant differences were found when using different doses, but different geometries had a greater effect. No differences were found between the threshold detectable calcification diameters at different heights in the breast. Calcification clusters may have a lower detectability using DBT than 2D imaging.


Assuntos
Doenças Mamárias/diagnóstico por imagem , Calcinose/diagnóstico por imagem , Mamografia/métodos , Imagens de Fantasmas , Intensificação de Imagem Radiográfica/métodos , Interpretação de Imagem Radiográfica Assistida por Computador/métodos , Tomografia por Raios X/métodos , Feminino , Humanos , Intensificação de Imagem Radiográfica/instrumentação , Software
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA