1.
Annu Int Conf IEEE Eng Med Biol Soc ; 2016: 3507-3510, 2016 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-28269054

RESUMO

The purpose of this study is to compare the performance of three nonlinear filters in online drift detection of continuous glucose monitors. The nonlinear filters are the extended Kalman filter (EKF), the unscented Kalman filter (UKF), and the particle filter (PF). They are all based on a nonlinear model of the glucose-insulin dynamics in people with type 1 diabetes. Drift is modelled by a Gaussian random walk and is detected based on the statistical tests of the 90-min prediction residuals of the filters. The unscented Kalman filter had the highest average F score of 85.9%, and the smallest average detection delay of 84.1%, with the average detection sensitivity of 82.6%, and average specificity of 91.0%.


Assuntos
Análise Química do Sangue/métodos , Glicemia/análise , Modelos Biológicos , Dinâmica não Linear , Análise Química do Sangue/instrumentação , Humanos , Distribuição Normal , Processamento de Sinais Assistido por Computador
SELEÇÃO DE REFERÊNCIAS