1.
Talanta ; 84(1): 34-41, 2011 Mar 15.
Artigo em Inglês | MEDLINE | ID: mdl-21315894

RESUMO

In this article a new coated platinum Cu(2+) ion selective electrode based on 2-((2-(2-(2-(2-hydroxy-5-methoxybenzylideneamino)phenyl)disufanyl)phenylimino) methyl)-4-methoxyphenol Schiff base (L(1)) as a new ionophore is described. This sensor has a wide linear range of concentration (1.2 × 10(-7)-1.0 × 10(-1) mol L(-1)) and a low detection limit of 9.8 × 10(-8) mol L(-1)of Cu(NO(3))(2). It has a Nernstian response with slope of 29.54 ± 1.62 mV decade(-1) and it is applicable in the pH range of 4.0-6.0 without any divergence in potential. The coated electrode has a short response time of approximately 9s and is stable at least for 3.5 months. The electrode shows a good selectivity for Cu(2+) ion toward a wide variety of metal ions. The proposed sensor was successfully applied for the determination of Cu(2+) ion in different real and environmental samples and as indicator electrode for potentiometric titration of Cu(2+) ion with EDTA.


Assuntos
Compostos de Benzil/química , Cobre/química , Bases de Schiff/química , Café/química , Cobre/análise , Eletrodos , Concentração de Íons de Hidrogênio , Limite de Detecção , Membranas Artificiais , Águas Minerais/análise , Platina/química , Cloreto de Polivinila/química , Potenciometria , Reprodutibilidade dos Testes , Rios/química , Solventes/química , Fatores de Tempo , Água/química