Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Thorac Cardiovasc Surg ; 114(3): 448-60, 1997 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-9305199

RESUMO

OBJECTIVE(S): The reuse of disposable devices is a potential source of significant cost savings to hospitals. Venous and arterial perfusion cannulas under new and reused conditions were selected to identify the clinical, safety, technical, logistic, and economic issues that must be addressed to realize these savings. METHODS: Single- and dual-stage venous and arterial cannulas from two manufacturers were tested when new, after initial clinical use, and after a single clinical use plus up to nine simulated reuses. Reuse was simulated by end-to-end bending, coupling and uncoupling of the connectors, and by two 1-hour soaks in plasma at 4 degrees and 40 degrees C, respectively. Cannulas were decontaminated and then processed by a peracetic acid-based liquid chemical sterilization system after each use/reuse. Sterilization was validated by eliminating Bacillus subtilis spores from the cannulas on each of five consecutive cycles. Cannulas were tested for physical changes, functional integrity, biocompatibility, and in vivo performance in sheep. A cost analysis was also performed. RESULTS: Sterilization was successfully achieved. Mechanical changes were less than 20% on all variables studied and were undetectable by experienced cardiac surgeons in selective evaluation. No clinically important differences were found between new and reused cannulas, even after nine simulated reuses. Reusing cannulas four times would reduce the cost per procedure from $53 to $19 (64%). CONCLUSIONS: Preliminary data suggest that the perfusion cannulas tested can be safely and efficaciously used five times. Limited reuse of these disposable cannulas is technically feasible and cost-effective. Cannula reuse would result in a small incremental savings; however, with more expensive devices and higher-volume sterilization procedures, the savings could be considerably greater. This program provides a model for evaluation of other single-use medical devices for reuse.


Assuntos
Cateterismo Periférico/instrumentação , Equipamentos Descartáveis , Animais , Bacillus subtilis , Materiais Biocompatíveis , Cateterismo Periférico/economia , Análise Custo-Benefício , Custos e Análise de Custo , Equipamentos Descartáveis/economia , Contaminação de Equipamentos , Reutilização de Equipamento/economia , Estudos de Viabilidade , Humanos , Teste de Materiais , Perfusão/instrumentação , Ovinos , Esterilização , Resistência à Tração , Anormalidade Torcional
2.
Biomed Instrum Technol ; 31(3): 248-9, 1997.
Artigo em Inglês | MEDLINE | ID: mdl-9181243

RESUMO

The reuse of disposable devices is a potential source of significant cost savings to hospitals. Venous and arterial perfusion cannulae under new and reused conditions were selected to identify the clinical, safety, technical, logistic, and economic issues that must be addressed to realize these savings. Single- and dual-stage venous and arterial cannulae from two manufacturers were tested when new, after initial clinical use, and after a single clinical use plus up to nine simulated reuses. Reuse was simulated by end-to-end bending, coupling and uncoupling the connectors, and by two 1-hour soaks in plasma at 4 degrees C and 40 degrees C, respectively. Cannulae were decontaminated and then sterilized by a peracetic acid based liquid chemical sterilization system following each use/reuse. Sterilization was validated by eliminating Bacillus subtilis spores from the cannulae on each of five consecutive cycles. Cannulae were tested for physical changes, functional integrity, biocompatibility, and in vivo performance in sheep. A cost minimization analysis was also performed. No clinically important differences were found between new and reused cannulae, even after nine simulated reuses. Mechanical changes were less than 20% on all variables studied and were undetectable by experienced cardiac surgeons in selective evaluation. Sterilization was successfully achieved. Reusing cannulae for times would reduce the cost per procedure from $53 to $19 (64%). Perfusion cannulae tested can be safely and efficaciously used five times. This study suggests that reuse would result in a small incremental savings; however, with more expensive devices and higher-volume sterilization procedures, the savings could be exponentially greater. Although this study demonstrates that it may be technically feasible and cost-effective to reuse disposable cannulae, the U.S. Food and Drug Administration does not sanction the reuse of disposable cannulae.


Assuntos
Cateterismo/instrumentação , Equipamentos Descartáveis , Perfusão/instrumentação , Animais , Bacillus subtilis/efeitos dos fármacos , Materiais Biocompatíveis , Procedimentos Cirúrgicos Cardíacos/instrumentação , Cateterismo/economia , Temperatura Baixa , Redução de Custos , Análise Custo-Benefício , Desinfetantes/uso terapêutico , Equipamentos Descartáveis/economia , Elasticidade , Desenho de Equipamento , Reutilização de Equipamento/economia , Segurança de Equipamentos , Estudos de Viabilidade , Custos Hospitalares , Temperatura Alta , Humanos , Ácido Peracético/uso terapêutico , Perfusão/economia , Plasma , Ovinos , Esporos Bacterianos/efeitos dos fármacos , Esterilização/métodos , Propriedades de Superfície , Estados Unidos , United States Food and Drug Administration
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...