1.
Pacing Clin Electrophysiol ; 34(9): 1087-91, 2011 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-21605137

RESUMO

BACKGROUND: Implantable heart rhythm devices are susceptible to interference in hospitals where electromagnetic interference (EMI) sources are ubiquitous. CASE DESCRIPTION: We report three cases in which EMI from the external defibrillator caused the inability to interrogate Boston Scientific cardiac resynchronization therapy-pacemaker (CRT-P) devices. We have documented interference with the Boston Scientific CRT-P Contact Renewal device model numbers H120/H125 (Natick, MA, USA) and two brands of external defibrillators: the Philips Heartstart XL model number M4735A (Andover, MA, USA) and the Hewlett-Packard Codemaster model number M1722B (Palo Alto, CA, USA). For device implants, we routinely place external pacing pads with the external defibrillator in the "standby" mode for transcutaneous pacing so that only the pacer "start/stop" button needs to be pressed when necessary. We have not been able to interrogate three consecutive Boston Scientific CRT-P devices prior to closure while the external defibrillator had the back-up pacing mode on "standby." In our initial case, a second device was opened because this interaction was not recognized. We documented EMI with the standby pacing mode ON and discovered that by disabling only the "standby" pacing mode on the external defibrillator, the device could be interrogated without difficulty. CONCLUSIONS: This is a case series reporting EMI with a Boston Scientific CRT-P Contact Renewal device H120/H125 telemetry from an external defibrillator with pacing mode on "standby." Failure to recognize this important interaction may lead to inappropriate device and resource utilization.


Assuntos
Dispositivos de Terapia de Ressincronização Cardíaca/efeitos adversos , Desfibriladores/efeitos adversos , Campos Eletromagnéticos/efeitos adversos , Análise de Falha de Equipamento
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...