Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Orthop Res ; 2024 Jul 02.
Artigo em Inglês | MEDLINE | ID: mdl-38953239

RESUMO

Resonance frequency analysis (RFA) is valuable for assessing implant status. In a previous investigation, acetabular cup fixation was assessed using laser RFA and the pull-down force was predicted in an in vitro setting. While the pull-down force alone is sufficient for initial fixation evaluation, it is desirable to evaluate the bone strength of the foundation for subsequent fixation. Diminished bone quality causes micromotion, migration, and protracted osseointegration, consequently elevating susceptibility to periprosthetic fractures and failure of ingrained trabecular bone. Limited research exists on the evaluation of bone mineral density (BMD) around the cup using RFA. For in vivo application of laser RFA, we implemented the sweep pulse excitation method and engineered an innovative laser RFA device having low laser energy and small dimensions. We focused on a specific frequency range (2500-4500 Hz), where the peak frequency was presumed to be influenced by foundational density. Quantitative computed tomography with a phantom was employed to assess periprosthetic BMD. Correlation between the resonance frequency within the designated range and the density around the cup was evaluated both in the laboratory and in vivo using the novel laser RFA device. The Kruskal-Wallis test showed robust correlations in both experiments (laboratory study: R = 0.728, p < 0.001; in vivo study: R = 0.619, p < 0.001). Our laser RFA system can assess the quality of bone surrounding the cup. Laser RFA holds promise in predicting the risk of loosening and might aid in the decision-making process for additional fixation through screw insertion.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...