We have reported two methods to analyze glucose in the interstitial fluid of skin based on mid-infrared excitation with a tunable quantum cascade laser and photoacoustic or photothermal detection. These methods were evaluated for optimum skin locations to obtain reproducible glucose information. The lower part of the arm, the hypothenar, the tips of the index finger and the thumb were tested. The thumb appears to be the optimal skin location, followed by the index finger. Basic requirements for an optimum site are good capillary blood perfusion, low Stratum corneum thickness and the absence of fat layers. To obtain a correlation on such a site, spectra were recorded on volunteers continuously after blood glucose manipulation. However, continuous measurements on an in vivo sample such as the skin have to cope with physiological alterations such as the formation of sweat. We have used both detection schemes to investigate the acid mantle reformation after washing during time scales similar to continuous measurements for calibration spectra. We found that reconstitution of the acid mantle of skin may be seen in less than one hour. Precleaning of the measurement site may thus be useful for intermittent, but not for long term continuous measurements.


Assuntos
Glucose/metabolismo , Técnicas Fotoacústicas/métodos , Pele/metabolismo , Espectrofotometria Infravermelho/métodos , Humanos , Masculino
2.
Spectrochim Acta A Mol Biomol Spectrosc ; 184: 220-227, 2017 Sep 05.
Artigo em Inglês | MEDLINE | ID: mdl-28500960

RESUMO

We have recently reported infrared spectroscopy of human skin in vivo using quantum cascade laser excitation and photoacoustic or photothermal detection for non-invasive glucose measurement . Here, we analyze the IR light diffusely reflected from skin layers for spectral contributions of glucose. Excitation of human skin by an external cavity tunable quantum cascade laser in the spectral region from 1000 to 1245cm-1, where glucose exhibits a fingerprint absorption, yields reflectance spectra with some contributions from glucose molecules. A simple three-layer model of skin was used to calculate the scattering intensities from the surface and from shallow and deeper layers using the Boltzmann radiation transfer equation. Backscattering of light at wavelengths around 10µm from the living skin occurs mostly from the Stratum corneum top layers and the shallow layers of the living epidermis. The analysis of the polarization of the backscattered light confirms this calculation. Polarization is essentially unchanged; only a very small fraction (<3%) is depolarized at 90° with respect to the laser polarization set at 0°. Based on these findings, we propose that the predominant part of the backscattered light is due to specular reflectance and to scattering from layers close to the surface. Diffusely reflected light from deeper layers undergoing one or more scattering processes would appear with significantly altered polarization. We thus conclude that a non-invasive glucose measurement based on backscattering of IR light from skin would have the drawback that only shallow layers containing some glucose at concentrations only weakly related to blood glucose are monitored.


Assuntos
Pele/química , Espectrofotometria Infravermelho/métodos , Líquido Extracelular/química , Glucose/análise , Glucose/química , Humanos , Lasers Semicondutores , Luz , Espalhamento de Radiação
3.
Analyst ; 142(3): 495-502, 2017 Jan 26.
Artigo em Inglês | MEDLINE | ID: mdl-28098265

RESUMO

An infrared spectroscopic technique is described that employs a mid-IR broadband (980-1245 cm-1) tunable quantum cascade laser (QCL) to produce a pump beam, and a detection method based on photothermal deflection, enhanced by total internal reflection. The IR spectra thus obtained are depth-dependent by modulating the pump beam with different frequencies between 10 Hz and 500 Hz. A model system consisting of glucose and a polymer film is used to demonstrate the depth selectivity of this technique. We also apply this photothermal depth profiling method to record in vivo IR spectra of the human epidermis at different depths. This information can be used for a non-invasive glucose monitoring on diabetes patients, which is also demonstrated. Beyond biomedical infrared spectroscopy, there are numerous applications for total internal reflection enhanced photothermal deflection spectroscopy (TIR-PTDS). The high penetration depth of mid-IR light compared to the traditional ATR-FTIR technique and the easy sample access make this technique appropriate for in situ measurements, such as in industrial quality control. The depth selectivity of TIR-PTDS may be a convincing argument for its use in the analysis of multilayered samples or for the analysis of artwork, where the layers of interest are covered by a layer of varnish.


Assuntos
Glicemia/análise , Pele/diagnóstico por imagem , Espectrofotometria Infravermelho , Glucose , Humanos
4.
Rev Sci Instrum ; 84(8): 084901, 2013 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-24007090

RESUMO

The application of a novel open, windowless cell for the photoacoustic infrared spectroscopy of human skin is described. This windowless cavity is tuned for optimum performance in the ultrasound range between 50 and 60 kHz. In combination with an external cavity tunable quantum cascade laser emitting in the range from ~1000 cm(-1) to 1245 cm(-1), this approach leads to high signal-to-noise-ratio (SNR) for mid-infrared spectra of human skin. This opens the possibility to measure in situ the absorption spectrum of human epidermis in the mid-infrared region at high SNR in a few (~5) seconds. Rapid measurement of skin spectra greatly reduces artifacts arising from movements. As compared to closed resonance cells, the windowless cell exhibits the advantage that the influence of air pressure variations, temperature changes, and air humidity buildup that are caused by the contact of the cell to the skin surface can be minimized. We demonstrate here that this approach can be used for continuous and non-invasive monitoring of the glucose level in human epidermis, and thus may form the basis for a non-invasive monitoring of the glucose level for diabetes patients.


Assuntos
Epiderme/química , Líquido Extracelular/química , Glucose/análise , Monitorização Fisiológica/instrumentação , Técnicas Fotoacústicas/instrumentação , Espectrofotometria Infravermelho/instrumentação , Ultrassom , Pressão do Ar , Teste de Tolerância a Glucose , Humanos , Umidade , Razão Sinal-Ruído , Temperatura
5.
Anal Chem ; 85(2): 1013-20, 2013 Jan 15.
Artigo em Inglês | MEDLINE | ID: mdl-23214424

RESUMO

The noninvasive determination of glucose in the interstitial layer of the human skin by mid-infrared spectroscopy is reported. The sensitivity for this measurement was obtained by combining the high pulse energy from an external cavity quantum cascade laser (EC-QCL) tunable in the infrared glucose fingerprint region (1000-1220 cm(-1)) focused on the skin, with a detection of the absorbance process by photoacoustic spectroscopy in the ultrasound region performed by a gas cell coupled to the skin. This combination facilitates a quantitative measurement for concentrations of skin glucose in the range from <50 mg/dL to >300 mg/dL, which is the relevant range for the glucose monitoring in diabetes patients. Since the interstitial fluid glucose level is representative of the blood glucose level and follows it without significant delay (<10 min), this method could be applied to establish a noninvasive, painless glucose measurement procedure that is urgently awaited by diabetes patients. We report here the design of the photoacoustic experiments, the spectroscopy of glucose in vivo, and the calibration method for the quantitative determination of glucose in skin. Finally, a preliminary test with healthy volunteers and volunteers suffering from diabetes mellitus demonstrates the viability of a noninvasive glucose monitoring for patients based on the combination of infrared QCL and photoacoustic detection.