Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Nucl Med ; 32(12): 2342-50, 1991 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-1744726

RESUMO

A standard set of performance measurements is proposed for use with positron emission tomographs. This set of measurements has been developed jointly by the Computer and Instrumentation Council of the Society of Nuclear Medicine and the National Electrical Manufacturers Association. The measurements include tests of spatial resolution, scatter fraction, sensitivity, count rate losses and randoms, uniformity, scatter correction, attenuation correction, and count rate linearity correction.


Assuntos
Tomografia Computadorizada de Emissão/normas , Humanos
2.
J Nucl Med ; 31(5): 610-6, 1990 May.
Artigo em Inglês | MEDLINE | ID: mdl-2341897

RESUMO

A high-resolution, whole-body positron camera, POSICAM 6.5 BGO, has been designed, built, and tested; results from it are presented. The camera utilizes 1,320 BGO crystals and 720 PMTS in a staggered geometry to produce high resolution of 5.8 mm FWHM and 21 image planes simultaneously. The axial resolution of the camera is measured at 11.9 mm at the center. High axial sampling is achieved with 5.125 mm separation of the image planes such that three-dimensional imaging of an object can be carried out in a single scan. Recovery of volumetric distribution of radioactivity and object dimensions in axial and sagittal views is demonstrated by imaging spherical objects 13 mm to 39 mm in diameter.


Assuntos
Tomografia Computadorizada de Emissão/instrumentação , Desenho de Equipamento , Estudos de Avaliação como Assunto
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...