Int J Prosthodont ; 28(5): 538-41, 2015.
Artigo em Inglês | MEDLINE | ID: mdl-26340017

RESUMO

The influence of repeated system-specific torque tightening on the position stability of the abutment after de- and reassembly of the implant components was evaluated in six dental implant systems with a conical implant-abutment connection. An established experimental setup was used in this study. Rotation, vertical displacement, and canting moments of the abutment were observed; they depended on the implant system (P = .001, P < .001, P = .006, respectively). Repeated torque tightening of the abutment screw does not eliminate changes in position of the abutment.


Assuntos
Dente Suporte , Projeto do Implante Dentário-Pivô , Retenção em Prótese Dentária , Fenômenos Biomecânicos , Humanos , Teste de Materiais , Rotação , Propriedades de Superfície , Torque
SELEÇÃO DE REFERÊNCIAS