Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Bone Joint Surg Br ; 93(12): 1597-601, 2011 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-22161920

RESUMO

This multicentre study analysed 12 alumina ceramic-on-ceramic components retrieved from squeaking total hip replacements after a mean of 23 months in situ (11 to 61). The rates and patterns of wear seen in these squeaking hips were compared with those seen in matched controls using retrieval data from 33 'silent' hip replacements with similar ceramic bearings. All 12 bearings showed evidence characteristic of edge-loading wear. The median rate of volumetric wear was 3.4 mm(3)/year for the acetabular component, 2.9 mm(3)/year on the femoral heads and 6.3 mm(3)/year for head and insert combined. This was up to 45 times greater than that of previously reported silent ceramic-on-ceramic retrievals. The rate of wear seen in ceramic components revised for squeaking hips appears to be much greater than in that seen in retrievals from 'silent' hips.


Assuntos
Artroplastia de Quadril/instrumentação , Análise de Falha de Equipamento/estatística & dados numéricos , Prótese de Quadril/efeitos adversos , Ruído , Complicações Pós-Operatórias/etiologia , Falha de Prótese/efeitos adversos , Adulto , Idoso , Óxido de Alumínio , Artroplastia de Quadril/efeitos adversos , Estudos de Casos e Controles , Cerâmica , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Desenho de Prótese/efeitos adversos , Falha de Tratamento