A powerful decoupling method is introduced to obtain decoupled signal voltages from quadrature coils in magnetic resonance imaging (MRI). The new method uses the knowledge of the position of the signal source in MRI, the active slice, to define a new mutual impedance which accurately quantifies the coupling voltages and enables them to be removed almost completely. Results show that by using the new decoupling method, the percentage errors in the decoupled voltages are of the order of 10(-7) % and isolations between two coils are more than 170 dB.


Assuntos
Aumento da Imagem/instrumentação , Interpretação de Imagem Assistida por Computador/instrumentação , Imageamento por Ressonância Magnética/instrumentação , Transdutores , Desenho Assistido por Computador , Desenho de Equipamento , Análise de Falha de Equipamento , Aumento da Imagem/métodos , Interpretação de Imagem Assistida por Computador/métodos , Reprodutibilidade dos Testes , Sensibilidade e Especificidade