1.
Phys Med Biol ; 56(18): 5877-88, 2011 Sep 21.
Artigo em Inglês | MEDLINE | ID: mdl-21852723

RESUMO

We present an initial evaluation of a mechanically cooled, high-purity germanium double-sided strip detector as a potential gamma camera for small-animal SPECT. It is 90 mm in diameter and 10 mm thick with two sets of 16 orthogonal strips that have a 4.5 mm width with a 5 mm pitch. We found an energy resolution of 0.96% at 140 keV, an intrinsic efficiency of 43.3% at 122 keV and a FWHM spatial resolution of approximately 1.5 mm. We demonstrated depth-of-interaction estimation capability through comparison of pinhole acquisitions with a point source on and off axes. Finally, a flood-corrected flood image exhibited a strip-level uniformity of less than 1%. This high-purity germanium offers many desirable properties for small-animal SPECT.


Assuntos
Germânio , Modelos Animais , Tomografia Computadorizada de Emissão de Fóton Único/instrumentação , Animais , Campos Eletromagnéticos , Desenho de Equipamento , Câmaras gama , Aumento da Imagem/instrumentação , Aumento da Imagem/métodos , Aceleradores de Partículas/instrumentação , Imagens de Fantasmas , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Tomografia Computadorizada de Emissão de Fóton Único/métodos