J Neurosci Methods ; 136(1): 99-102, 2004 Jun 15.
Artigo em Inglês | MEDLINE | ID: mdl-15126050

RESUMO

Typically small animal radiological images are obtained after placing the animal in the center of the imaging device using beds or platforms, and then adjusting the position after obtaining a scout image. Such a process does not permit the reproducible visualization of the same anatomical plane with repeated examinations. We have developed a device that allows stereotaxic placement of an animal in precisely the same position for repeated examinations. The instrument incorporates a full range of physiological monitoring and life support systems including temperature control, anesthesia delivery and respiratory monitoring. Using magnetic resonance imaging (MRI), the accuracy and reliability of this device is demonstrated in a rat traumatic brain injury (TBI) model.


Assuntos
Lesões Encefálicas/patologia , Encéfalo/patologia , Imagem de Difusão por Ressonância Magnética/instrumentação , Imagem de Difusão por Ressonância Magnética/métodos , Técnicas Estereotáxicas/instrumentação , Animais , Masculino , Ratos , Ratos Sprague-Dawley , Reprodutibilidade dos Testes
SELEÇÃO DE REFERÊNCIAS