Pacing Clin Electrophysiol ; 26(5): 1245-8, 2003 May.
Artigo em Inglês | MEDLINE | ID: mdl-12765453

RESUMO

Transvenous temporary pacing is associated with a substantial dislocation rate reported to range from 10 to 37%. The aim of the study was to assess the safety and utility of a recently introduced 3.5 Fr temporary pacing lead using active fixation in a consecutive series of 36 patients with prolonged (> or = 48 hours) transvenous temporary pacing (validation group). A group of 36 patients with prolonged transvenous pacing managed with a passive-fixation lead just prior to the introduction of the active-fixation lead served as a control group (reference group). Pacing related adverse events included dislocation, inappropriate pacing (i.e., two-fold or greater increase of initial pacing threshold), local infection, and thrombosis. There were no significant differences in patient characteristics or duration of pacing (5.84 +/- 2.4 days in the reference group vs 5.94 +/- 2.6 days in the validation group). Acute pacing threshold was significantly higher in the validation group as compared to the reference group (1.38 +/- 0.67 V vs. 0.7 +/- 0.21 V, P < 0.01). The dislocation rate was significantly lower in the validation group as compared to the reference group (5.5 vs 33.3%, P < 0.001). There were 11 (31%) pacing related adverse events in the validation group versus 21 (58%) in the reference group (P < 0.01). The vast majority of patients in the validation group (75%) had ambulatory temporary pacing. Thus, transvenous temporary pacing using active fixation is safe and is associated with a low dislocation rate and a reduction in pacing related adverse events.


Assuntos
Estimulação Cardíaca Artificial/métodos , Marca-Passo Artificial , Idoso , Bradicardia/terapia , Estimulação Cardíaca Artificial/efeitos adversos , Falha de Equipamento , Feminino , Veia Femoral , Parada Cardíaca/terapia , Humanos , Masculino , Marca-Passo Artificial/efeitos adversos , Estatísticas não Paramétricas
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...