Int J Audiol ; 49(10): 775-87, 2010 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-20666693

RESUMO

The Nucleus CI24RE 'Freedom' device offers higher stimulation rates and lower noise levels in action potential measurements (ECAPs) than previous devices. A study including ten European implant teams showed that the effect of changes in rate from 250 to 3500 pulses per second on tilt and curvature of the T and C profiles is insignificant. When changing rate one may change the levels at all electrodes by the same amount. Using an automated procedure ECAPs could be measured quickly and reliably at a noise level of only 1 microV, this did not result in improved correlations between the tilt and curvature parameters of the ECAP profiles and those of the T and C profiles. Average C levels appear to differ markedly among implant centers; a better assessment protocol is required. When increasing stimulus rate one should take into account that this requires higher pulse charges per second and more power consumption.


Assuntos
Potenciais de Ação , Percepção Auditiva , Limiar Auditivo , Implantes Cocleares , Percepção Sonora , Adulto , Idoso , Automação , Limiar Diferencial , Estimulação Elétrica/métodos , Humanos , Pessoa de Meia-Idade , Ruído , Análise de Componente Principal , Adulto Jovem