Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Nucl Med Commun ; 38(10): 837-842, 2017 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-28777221

RESUMO

OBJECTIVE: The aim of this study was to evaluate angular, spatial, and energy resolution, sensitivity, and shielding of a gamma-probe. MATERIALS AND METHODS: The EUROPROBE II gamma-probe (EuroRad) with sources of technetium-99m was assessed according to NEMA NU-3-2004. Resolution tests were evaluated considering the full width at half maximum (FWHM). The following parameters were evaluated: angular resolution in air, spatial resolution with a scattering medium and in air, energy resolution, and sensitivity and shielding. The collimator was used to evaluate angular and spatial resolution, sensitivity, and shielding. Background radiation was considered and did not affect the counts. RESULTS: FWHM of angular resolution (at 3/30 cm) was 39.17°/33.13° with the collimator and 74.08°/71.51° without the collimator; FWHM of spatial resolution in air at 10 mm was 13.32 mm with the collimator and 21.23 mm without the collimator. Energy resolution (%FWHM) was 20.51%. Sensitivity at 10 mm was 4.642±5 cps/MBq without the collimator and 1.063±2 cps/MBq with the collimator; shielding effectiveness of the probe tip was 99.52%. Background was not relevant to the counts. CONCLUSION: We showed that the collimator improved angular and spatial resolution to the detriment of sensitivity. Feasible results of energy resolution, sensitivity, and shielding were achieved.


Assuntos
Raios gama , Proteção Radiológica , Biópsia de Linfonodo Sentinela/instrumentação , Raios gama/efeitos adversos , Biópsia de Linfonodo Sentinela/efeitos adversos , Tecnécio/efeitos adversos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...