Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Assunto principal
Intervalo de ano de publicação
1.
Anal Bioanal Chem ; 381(3): 713-20, 2005 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-15696276

RESUMO

Techniques that allow the elements present in the air to be determined in a simple and rapid manner are very attractive. Direct aerosol sampling techniques avoid the need to pretreat the filter via wet digestion in order to remove any sources of contamination, and they decrease the precipitation time significantly. Analyzers based on this technique can also determine the concentrations of elements in the air automatically in situ. This paper is concerned with the development of a novel analytical system that is based on electrostatically precipitating aerosols from the air into a graphite furnace. The equipment includes a Zeeman spectrometer with high frequency modulation polarization (MGA-915), and an electrostatic precipitation system incorporated into the analyzer. The high sensitivity of the system developed here means that it can be used to determine element concentrations in the air exhaled by humans, as well as those in ambient air.


Assuntos
Ar/análise , Expiração , Aerossóis , Desenho de Equipamento , Humanos , Reprodutibilidade dos Testes , Análise Espectral/instrumentação , Análise Espectral/métodos , Eletricidade Estática