In Vitro Cell Dev Biol Anim ; 35(2): 87-93, 1999 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-10475262

RESUMO

In the present study, a unique mechanical strain apparatus for three-dimensional organotypic cultures was developed into a computerized system. It consists of a personal computer running Windows-based software, the Bio-Stretch Manager, a Bio-Stretch Controller, and three sets of magnet boards. Cells are cultured on a Gelfoam sponge that is placed in a 35 mm petri dish with one end glued to the dish, and the other end attached to a coated steel bar. The petri dish is placed in front of a magnet, and the movement of the steel bar is controlled by dynamically changing the magnetic field. Up to five stretch patterns of variable frequency, duty cycle, and magnitude can be designed for each stretch regimen. Three different stretch regimens can be tested simultaneously. The operational characteristics of sponges were examined. Attachment of cells to the sponges was observed on several cell types. These features provide wide options for using this system to study the effects of mechanical stretch on cells.


Assuntos
Biologia Celular/instrumentação , Técnicas de Cultura de Células , Fenômenos Fisiológicos Celulares , Estresse Mecânico , Animais , Aorta , Computadores , Endotélio Vascular/citologia , Esponja de Gelatina Absorvível , Músculo Liso Vascular/citologia , Miocárdio/citologia , Ratos , Reprodutibilidade dos Testes , Sensibilidade e Especificidade