Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Acta Obstet Gynecol Scand ; 101(2): 241-247, 2022 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-35049047

RESUMO

INTRODUCTION: In this feasibility study, we hypothesize that the evaluation of cervical biomechanical strength can be improved if cervical length measurement is supplemented with quantitative elastography, which is a technique based on conventional ultrasound elastography combined with a force-measuring device. Our aims were to: (a) develop a force-measuring device; (b) introduce a cervical elastography index (CEI) and a cervical strength index (CSI; defined as cervical length × CEI); (c) evaluate how these indexes assess the cervical softening that takes place during normal pregnancy; and (d) how these indexes predict the cervical dilatation time from 4 to 10 cm. MATERIAL AND METHODS: An electronic force-measuring device was mounted on the handle of the transvaginal probe, allowing for force measurement when conducting elastography. The study group concerned with normal cervical softening included 44 unselected pregnant women. Outcomes were CEI and CSI at different gestational ages. The study group for labor induction included 26 singleton term pregnant women admitted for labor induction. Outcome was defined as cervical dilatation time from 4 to 10 cm. Elastography measured the changes in mean gray value (intensity) during manual compressions. Region of interest was set within the anterior cervical lip. RESULTS: We found that the mean of all variables regarding cervical softening decreased from early to late pregnancy: ie cervical length from 34 to 29 mm, CEI from 0.17 to 0.11 N, and CSI from 5.9 to 3.1 N mm. Moreover, the cervical dilatation time during labor induction was associated with CEI, although not statistically significantly (area under the ROC curve of 0.67), but not with the Bishop score, the cervical length, or the CSI. CONCLUSIONS: We propose that quantitative elastography based on changes in the intensity of the B-mode ultrasound recording, in combination with a force-measuring device on the handle of the vaginal probe, deserves further investigation as an approach for evaluation of cervical biomechanical strength.


Assuntos
Colo do Útero/fisiologia , Técnicas de Imagem por Elasticidade/instrumentação , Ultrassonografia Pré-Natal , Adolescente , Adulto , Colo do Útero/diagnóstico por imagem , Desenho de Equipamento , Feminino , Idade Gestacional , Humanos , Recém-Nascido , Valor Preditivo dos Testes , Gravidez , Trimestres da Gravidez , Nascimento Prematuro , Adulto Jovem
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...