Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Masui ; 53(11): 1315-6, 2004 Nov.
Artigo em Japonês | MEDLINE | ID: mdl-15587190

RESUMO

Terumo's Surshield Surflow II i.v. catheter automatically engages a stainless steel clip to shield its needle tip when the needle goes out from the catheter hub. However, in our experience the safety clip of Surshield Surflow II remains in the catheter hub in a high proportion of cases when the catheter hub is held with a pair of forceps. The hub wall is so thin that the force existed to hold the hub can be easily transmitted to the safety clip. Another factor is the loose attachment of the safety clip to the needle tip. To prevent needle stick injury, further improvement of safety i.v. catheters is necessary to lead their increased use.


Assuntos
Cateterismo/instrumentação , Cateteres de Demora , Ferimentos Penetrantes Produzidos por Agulha/prevenção & controle , Equipamentos de Proteção , Desenho de Equipamento , Segurança de Equipamentos , Humanos , Agulhas
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...